Use the VMS-1655M Vortex Monitor in the Flowsafe Stable Vortex II fume hood and Flowsafe Stable Vortex conversion kit to measure
and control the vortex pressure and adjust to maintain correct fume hood containment. The VMS-1655M adjusts for environmental
and usage conditions such as sash movement, fume hood loading, room pressure fluctuations, and cross drafts with active vortex
stabilization in real-time. The alarm function notifies you if there is a compromise in fume hood performance. Some of the features and
benefits include:
•
Full-color touchscreen display with programmable display options and adjustable back-light
•
360° Safety Halo
™
•
Intuitive user interface that simplifies setup and configuration of unit
•
Optional display mode to show real-time face velocity at main screen
•
Graphical hood status display
•
Audible and visual alarms
•
Multi-level password protection
•
1 analog input: 0 VDC to 10 VDC
•
2 analog outputs, actuator control, exhaust air flow feedback
•
BACnet
® MS/TP network comms
•
Real-time view facility facilitates diagnostics during setup and commissioning
Overview
IMPORTANT:
Make all wiring connections in accordance with the National Electrical Code and local regulations. Use proper
Electrostatic Discharge (ESD) precautions during installation and servicing to avoid damaging the electronic circuits of the
VMS-
1655M Fume Hood Monitor
.
IMPORTANT :
Assurez-vous que tous les branchements de câbles sont effectués selon le Code national de l’électricité et les
réglementations locales. Utilisez une bonne protection contre les décharges électrostatiques (ESD) pendant l’installation et l’entretien
pour éviter d’endommager les circuits électroniques du
VMS-1655M Fume Hood Monitor
.