Page 3

OPERATING INSTRUCTIONS

Adjust Height

--To raise or lower the mat table, simply touch the up or down arrow key on the hand switch.  The

low voltage hand switch has operating power of VDC/mA of 12/50.

Duty Cycle

--System is an intermittent operation; 1 minute on, 9 minutes off.

Capacity

--The table is intended to support 

750 lbs

(333.6 Kilograms).

Casters

--Locking the casters is accomplished by pushing down or stepping on the back part of the brake pedal

firmly.  To release the brake pedal, push down or step on the upper part.

Intended Use

--It is a powered table intended for medical purposes (Class I Device) that is an electrically oper-

ated flat surface table that can be adjusted to various positions.  It is used by patients with circulatory, neuro-
logical, or musculoskeletal conditions to increase tolerance to an upright or sitting position; [21CFR890.3760].
At no time should patients be left unattended on the mat table.

Intended  User  Profile

--It  is  intended  to  be  used  by  Licensed  Physical  Therapists,  Assistant  Physical

Therapists, Occupational Therapists, Assistant Occupational Therapists, and licensed rehab specialists utilizing
a prescribed therapeutic protocol, written by a licensed medical practitioner  or physician.  It is not intended to
be used by maintenance staff, installers, patients and/or lay persons.

MAINTENANCE

The mat table is equipped with a strong high-performance quiet linear actuator, having a long lifetime and vir-
tually  maintenance  free;  however,  conducting  routine  preventative  maintenance  checkups  are  a  must.    One
must routinely check the actuator for wear on the end mounts and connecting bolts, for unusual noises, for actu-
ator wobbling when extending or retracting the spindle, and that it is operating smoothly, as minimums.  Due to
normal wear, it may be necessary to replace the linear actuator.  The actuator is non-servicable in the field.  Call
customer service for replacement actuator.

The gas springs are used to assist the linear actuator in the first stages of the lifting process and may need
replacing from time to time, due to normal wear.  If one should notice an oil leak or oil spot on the floor below
the gas spring, or if table seems to be deficient in power, it may indicate the spring needs replacing.  If opera-
tor hears an unusual noise while operating the device, it may be coming from the gas spring, which would indi-
cate it may need replacing.  Call customer service to assist in interpreting the nature of the noise.

Due to normal wear, it may be necessary to replace the hand switch or the power cord.  Call our customer serv-
ice for replacement parts and/or installation instructions.

Cleaning the vinyl cover is accomplished with a mild detergent and water.  Do not use Peroxide, Hylex, Lysol,
or strong cleaning agents as they will discolor and reduce the life of the vinyl.  We also recommend diluting
some of the well known over the counter cleaners available before using; such as 409

cleaner with a ratio of

4:1 at a minimum.  A soft dust cloth will work to remove dust and lint on the steel frame when necessary.

Any repair should be performed by trained or certified service personnel.

Содержание TG1800 Series

Страница 1: ...d the information listed below for quick accessibility Model No _________________ Serial No _________________ Purchase Date _____________ Purchased From __________________________ Tri W G Service 1 800 437 8011 USER MANUAL P N 040115 750 For HI LO MAT TABLE TG1800 SERIES with Instructions for Use Read this user manual carefully before using Hi Lo Mat Table ...

Страница 2: ...Tri W G Inc ...

Страница 3: ... 2 SAFETY PRECAUTIONARY INSTRUCTIONS 2 OPERATING INSTRUCTIONS 3 MAINTENANCE 3 SYMBOLS INDICATORS and DEGREE OF SAFETY 4 WARNINGS and USAGE HINTS 4 OTHER HAZARDS 4 TROUBLESHOOTING 5 SPECIFICATIONS 5 NOTICE 5 WARRANTY 6 DISCLAIMER 7 TABLE OF CONTENTS ...

Страница 4: ...having flat level floors so table is not distorted out of alignment and caster brakes will function as intended 3 Casters should be locked at all times except for when moving Casters should be locked when a person is on the table 4 Connect power cord to properly grounded 120 Volt AC receptacle conforming to applicable nationally rec ognized electrical codes 5 Power supply cord serves as disconnect...

Страница 5: ...wever conducting routine preventative maintenance checkups are a must One must routinely check the actuator for wear on the end mounts and connecting bolts for unusual noises for actu ator wobbling when extending or retracting the spindle and that it is operating smoothly as minimums Due to normal wear it may be necessary to replace the linear actuator The actuator is non servicable in the field C...

Страница 6: ...ructively and with protective measures minimized the effects of existing hazards Pay attention to the following hazards and the possible countermeasures given in the following Attention Consult accompanying documents Protection against harmful ingress of water Ordinary Not suitable for use in the presence of a flammable anesthetic mixture with air or oxygen or nitrous oxide Input 120 Volt AC 60 Hz...

Страница 7: ...steel Finish Polyurethane Pearl Grey standard Porcelain optional Accessories Sit to Stand Guard Rails Various Exercise Accessories available Switch Handheld height adjusting switch low voltage VDC mA of 12 50 Power supply serves as main discon nect device Electrical 6 5A 120V 780W hospital grade plug NOTICE Design changes may have occurred in the product since this manual was published The technic...

Страница 8: ... or remanufactured Parts when the defective Part or Parts are returned to Tri W G Inc facilities at Buyer s expense upon 20 days prior written notice to Tri W G Inc Buyer will be charged for any replacement Parts when shipped to Buyer by Tri W G Inc When the defective Parts are returned to Tri W G Inc pursuant to Tri W G Inc returned goods authorization charges will be waived This limited warranty...

Страница 9: ...ormation contained herein The label s on any of the products components supercedes the copy and or data that is in this User Manual Copy data and or print on serial tags and labels attached to any of Tri W G products and or components supplied by other vendors used in Tri W G products shall supersede any and all copy data and or print stated in this User Manual However because of continuous progre...

Страница 10: ...Page 8 This page intended to not be used ...

Страница 11: ...Page 9 This page intended to not be used ...

Страница 12: ...Copyright 2015 Tri W G Inc All Rights Reserved Phone 1 800 437 8011 Email triwg triwg com 215 12th Avenue NE PO Box 905 Valley City ND 58072 0905 Manufactured and Distributed by ...

Отзывы: