LABORATORY REFERENCE STANDARDS
3000 PRECISION SERIES
Ü
3000RS
PRECISION RESISTANCE STANDARD
Страница 1: ...LABORATORY REFERENCE STANDARDS 3000 PRECISION SERIES 3000RS PRECISION RESISTANCE STANDARD...
Страница 2: ...Version 1 40 September 2011 All product names are trademarks of their respective companies 3000RS Series Precision Resistance Standard Operation Manual...
Страница 3: ...ed by the purchaser Transmille will contact the purchaser before proceeding with any repair To obtain repair under this guarantee the purchaser must send the instrument in its original packaging carri...
Страница 4: ...ransmille Ltd Page 3 TABLE OF CONTENTS GUARANTEE AND SERVICE 2 3000RS PRECISION RESISTANCE STANDARD 4 MAIN FEATURES 4 DESIGN NOTES 5 FEATURES 5 OPERATING NOTES 6 CONNECTING TO THE 3000RS 6 CONNECTION...
Страница 5: ...nge of multi product calibrators The stability and temperature co efficients of the 3000RS is designed for a working environment of 18 23 C and the standard is supplied complete with a calibration cer...
Страница 6: ...high precision wire wound for the 1 10 MOhm no adjustable resistors are used within the design of this reference Features Functions of the terminal post connections 3000RS Terminal Post Function CURRE...
Страница 7: ...tal considerations for the 3000RS precision resistance standard Follow these instructions when operating or storing the resistance standard Connecting to the 3000RS CAUTION Do not apply voltages above...
Страница 8: ...ow For the case to act as a shield the EARTH terminal must be connected to earth ground or instrument guard at some point in a calibration configuration When connecting two or more instruments with gu...
Страница 9: ...3000RS PRECISION RESISTANCE STANDARD OPERATION MANUAL Transmille Ltd Page 8 Fig 2 2 Wire Connection Example Fig 3 Earth Connection Example...
Страница 10: ...erence OUTPUTS Decade resistances from 1 Ohm to 10MOhm separate terminals for each resistor DIMENSIONS 45cm x 44cm x 9cm PERFORMANCE SPECIFICATIONS VALUE T C ppm C STABILITY YEAR ppm CONNECTION 1 Ohm...
Страница 11: ...t is recommended CAUTION Do not use cleaning fluids or solvents for cleaning as these may damage the enclosure and affect the plastic materials used in the precision resistance standard Handling Preca...