Power
Depressing the Power Button [B] turns on the orange 

LEDs. Depressing the button again, adds the red LEDs to 

the orange LEDs. Depressing the button again turn the 

unit off.
Depressing the Brightness Button [C] for one second will 

turn the unit off.
The LEDs will automatically shut off after 7 minutes.

Brightness Control
The Veinlite PEDI2 has three levels of brightness that are 

selected by pressing the Brightness Button [C] after the 

unit is turned on. Brightness control is important when 

working with small children.

Disposable Plastic Cover
The single-use disposable plastic covers for 

the Veinlite PEDI2 (Product #VLP2-DPC) 

MUST be used when the device is placed in 

contact with a patient’s skin.
CAUTION: These plastic covers are not sterile. They 

should be wiped with a germicidal wipe prior to use.
Before using Veinlite PEDI2, place it inside the disposable 

cover with the flap on the same side as the LEDs.
Pull the flap over the device to the other side.
Wipe the LED side of the Veinlite PEDI2 and any part of it 

coming in contact with the patient, with a medically 

approved germicidal wipe.

Batteries
The Veinlite PEDI2 is powered by 2 AAA batteries. They 

provide between 2 and 4 hours of continuous illumination 

depending on the number of LEDs turned on and the 

brightness level used.

Low Battery Indicator
When the batteries get low, the orange lights will flicker 

when turned on. Replace the batteries as soon as the 

orange light flickers when turned on.

Installing Or Replacing Batteries 
To install or replace the batteries, press firmly on the 

Battery Cover [D] and slide it off. Remove any used 

batteries. Install 2 new AAA batteries, oriented as 

indicated by the markings inside the battery compartment 

then slide the battery cover back into place.

Description
The PEDI2 utilizes the patented

1

 transillumination 

technique known as side-transillumination and a 

patented

2

 orange color LED light. A ring of bright light 

from 10 orange and 4 red LEDs is aimed through a clear 

plastic LED Window [A] and focused below the skin 

surface and towards the ring’s center.
The focused light creates a volume of illunimation, with 

the central focus acting as a virtual light source under the 

skin. This technique achieves unifrom illumination of a 

small region of tissue anywhere on the body.
The patented orange light is selected because of high 

absorption in venus blood. The red LEDs have high 

penetration in darkly pigmented skins. The combination of 

the two provides optimum imaging of veins in humans.

Vein Access
1
  Place the Veinlite PEDI2 inside the disposable plastic 

cover so that the device is totally covered.

2  Place the Veinlite PEDI2 on the skin, with the LED 

Window [A] facing down. Apply gentle pressure to the 

Veinlite PEDI2 to ensure it is in contact with the skin.

3  Depress the Power Button [B] to turn on the orange 

and then the red LEDs. Use the Brightness Button [C

to control the brightness level of the device to the 

desired level.

4  To scan for veins, place 

Veinlite PEDI2 against skin 

perpendicular to vein 

direction. Scan it 

forward and 

backward. Veins will be 

visible as dark lines due to 

transillumination.

5  Once a suitable vein 

has been identified, 

rotate device 90˚, 

parallel to vein.

6  Apply gentle pressure on the 

Veinlite PEDI2 to secure the 

vein.

7  Push back slightly on the 

Veinlite PEDI2 to stretch the 

skin and provide traction.

8  Insert the needle into 

the vein, through the 

opening of the 

device.

  NOTE: Discard the 

single-use disposable plastic 

cover after each patient.

READ INSTRUCTIONS BEFORE USE

TransLite LLC

345 Commerce Green Blvd

Sugar Land TX 77478 USA

+1 281 240 3111

www.veinlite.com

[email protected]

Caution

  Veinlite PEDI2 is designed for external examination 

only.

  To avoid contamination of the device, Veinlite PEDI2 

must be used with the disposable plastic cover.

  To avoid the spread of infection, always change the 

disposable plastic cover between patients and wipe 

the new cover with a germicidal wipe.

  These devices are intended for use only by clinicians 

trained in performing vascular access, or under the 

supervision of a clinician.

  Do not look directly into the LEDs while on.

  Device may be wiped down with a germicidal wipe.

  Do not immerse the device in liquid.

  Do not autoclave the device.

Introduction
The TransLite Veinlite PEDI2 is a hand-held 

battery-powered transillumination device for visualizing 

veins and superficial blood vessels in pediatric patients.

      

Veinlite 

PEDI2

LED Window [A]

Battery Cover [D]

Power Button [B]

Brightness Button [C]

Warranty
The Veinlite PEDI2 has a one year parts and labor 

warranty. It is not necessary to register your Veinlite PEDI2 

for warranty. The process is automatic and the serial 

number of the unit provides all the information needed.

Troubleshooting
Before returning the device to TransLite for repair, please 

call or e-mail to obtain an authorization number and 

shipping details.
No refunds will be given for devices shipped without 

authorization.

Specifications
Number of LEDs: 

10 Orange

  

4 Red

Weight with battery: 

29g

Dimensions: 

40.8×86.6×21mm

Battery: 

2x AAA 1.5V

Intended Use/Indications

  Emergency venous access in infants and children.

  Venous access in infants and pediatric patients with a 

history of difficult venous access.

  Easier visualization of veins in dark 

pigmented patients.

  Easier visualization of veins in overweight 

infants and children.

Contraindications

  When used for internal examination

  When used without the disposable plastic cover

Explanation of Symbols
 

Model Number

 

Serial Number

 Manufacturer

 

EC Representative (EU Only)

 

Do not dispose of as unsorted municipal waste

M. Devices Group

Health Education Centre

The Church, Portland Street

Southport PR8 1HU

United Kingdom

+44 1704 544 944

5

6

7

8

4

Отзывы: