Follow these steps to clean the accessories.
1.
Fully immerse the accessory in the cleaning solution. DO NOT immerse the
Transcend Micro device.
NOTE
:
HME Cartridge must be removed prior to cleaning the HME Adapter. The HME
cartridge cannot be cleaned; discard after 3-7 days of use.
2.
While immersed, thoroughly wipe the surface with a lint-free cotton cloth. Apply
firm pressure and ensure contact with all accessible connection surfaces to
adequately remove soil buildup.
3.
Clean the inside of the accessory by lifting then lowering the ends of the accessory,
while the accessory is filled with cleaning solution.
4.
Rinse accessory by immersing in distilled water. Move the accessory in a back-and-
forth motion for approximately 10 seconds to remove cleaning agent residue.
5.
Dry the outside of the accessory with a dry, lint-free cotton cloth. Allow the
accessory to air dry until the inside is dry. Length of drying time will depend on
ambient conditions.
Transcend batteries
•
DO NOT immerse batteries in water or cleaning solution.
•
For long term storage, which is any duration of time greater than three months, the
battery should be:
o
Stored at a temperature above -20C/-4F but below 20C/68F.
o
Stored in a dry location and with an approximate 30% charge. Allow 2 hours
to charge battery when fully depleted to achieve 30% charge
See the Transcend PowerAway User Guide on myTranscend.com for guidance information
on setup, use and caring for Transcend batteries.
Replacing the Transcend Micro PureFresh™
air filter
The Transcend Micro PureFresh air filter should be replaced at a minimum every six
months.