Transcend 365, Transcend 3 and Transcend 2 Troubleshooting Guide 
 

400287 Rev A 
 

Page | 11 

 

PAP Shutting Down 

 

If the CPAP is shutting down by its self in the middle of therapy, this is most likely caused by excessive 
leak and mask seal.  Check the following: 

 

Mask fitting 

 

UHA is properly attached (Transcend 2 only).  Remove and replace securely 

 

Check hose is properly attached to the UHA and there are no holes in the hose 

 

Download PAP compliance to verify leak rate (an acceptable leak rate is <10) 

 

If problem persists, contact Somnetics Customer Service 

 

PAP Blows Excessively (and Shuts Down)

 

The most common reason for this issue is due to the UHA (Transcend 2) not being connected or 
improperly connected.   This can also be caused by a dislodged mask during therapy, using a nasal mask 
when primarily breathing through mouth or having an incorrect seating of the mask. Download 

compliance report to verify leak rate isn’t the issue. (an acceptable leak rate is <10).  If leak is not the 

issue, contact Somnetics Customer Service. 

 

P8 Battery 

Non-functioning LED Light

 

The customer should test the battery function by charging it for 8 hours and then using it until the 
battery dies.  This will tell us if it is a functional issue or just the LED light is no longer functioning. 

 

Battery Usage

 

The P8 battery should last approximately 14 hours with the pressure of 14cmH2o on a full charge. Leak, 
altitude, breathing rate, and pressure can affect the time a battery will last. If the battery is not lasting 
as long as it should, check the compliance data to see if a leak is causing the battery to drain faster. If no 
leak present and settings seem within range, contact Somnetics Customer Service. 

 
Battery Life

 

1.

 

Expect roughly a 3-5% annual self-discharge during storage. This means that the battery will 
gradually drain when stored; however, it is not significant enough to require recharge for 
normal storage. 

2.

 

To maximize battery life when putting the battery into long term storage (3 months or greater); 

 

Store at a temperature below 20C/68F but greater than -20C/-4F 

Содержание 2

Страница 1: ... s must be submitted by email to RMA somnetics com with the above information included Quick Find Transcend 365 Transcend 3 and Transcend 2 Flashing Yellow Droplet and Yellow Diamond Key Notes Flashing Blue Droplet Fault 3 Flashing Red Droplet and X Fault 4 Flashing Red Fault 7 Flashing Red X and Heater Error Fault 9 Fault 1 Fault 11 Fault 3 Fault 12 Fault 4 Flashing Green LED Fault 7 No LED Light...

Страница 2: ...clean all surfaces 3 Replace the zero water filter cartridge if applicable 4 Run the drying mode on the CPAP and afterwards remove all power to reset the device 5 Fill the reservoir with distilled water and attach to CPAP 6 Attempt to run the device again If yellow droplet persists you should let the device run for one night as some residual non distilled water may still be present in the heater w...

Страница 3: ...temperature cutoff condition and turned off Causes Water signal failed to detect empty reservoir Caused by moisture getting inside the CPAP Poor wicking Poor contact between cellulose wick and heater ceramic Troubleshooting This error can correct itself on the next therapy if not 1 If applicable replace cellulose wick 2 Remove and drain the water reservoir 3 Run the drying mode on the CPAP and aft...

Страница 4: ...ustomer Service Fault 1 Issue Date Time have been reset Causes Low No voltage on clock battery Loose broken battery solder joint Troubleshooting 1 Check customer firmware if 2 1 2 or later there is an issue with the battery 2 Set up RMA Fault 3 Issue Pressure sensor measured a pressure that was above operating ranges Causes Issue in the air pathway Faulty pressure sensor or electronics Pinched or ...

Страница 5: ... the error persists Fault 7 Issue Pressure sensor ambient reading out of range Causes Extreme temperature changes Leaks in the air pathway Faulty pressure sensor or electronics Pinched or disconnected sensor tube Troubleshooting 1 If the device has been exposed to an extreme temperature hot or cold remove power and allow the PAP adjust to room temperature Then apply power and let the PAP sit in st...

Страница 6: ...in the CPAP Faulty blower motor or lose connections Troubleshooting 1 Run the drying mode on the CPAP and afterwards remove all power to reset the device 2 Attempt to run the device again 3 Set up an RMA if the error persists Fault 12 Issue Stalled Blower and Low Power Detected Causes Insufficient power to drive the blower Troubleshooting 1 Ensure adequate power supply 2 If using the battery plug ...

Страница 7: ...ate Confirm device is plugged into an active power supply outlet is best for testing and turn the device on Note The device will automatically turn off the LCD after 30 seconds of idle time no button presses press any button to re illuminate the LCD in this instance If issues persist contact Somnetics Customer Service No Airflow from the Transcend 365 miniCPAP Contact Somnetics Customer Service Wa...

Страница 8: ... start over again This will give the error code Fault Codes Fault 3 Issue Pressure sensor measured a pressure that was above operating ranges Causes Issue in the air pathway Faulty pressure sensor or electronics Pinched or disconnected sensor tube Troubleshooting 1 Reset the device 2 Check all connections and fittings to make sure there are no issues such as blockages 3 Attempt to run the device a...

Страница 9: ...or disconnected sensor tube Troubleshooting 1 If the device has been exposed to an extreme temperature hot or cold remove power and allow the PAP adjust to room temperature Then apply power and let the PAP sit in standby no blower for at least 20 minutes to allow it to obtain a new ambient pressure reading 2 Reset the device 3 Check all connections and fittings to make sure there are no issues 4 A...

Страница 10: ...eset the device and attempt therapy again 4 Set up an RMA if the error persists Issues Not Faults Flashing Green LED This means the filter needs to cleaned or replaced Once cleaned or replaced reset the flashing green LED by holding down the power button for 15 seconds No LED Lights Unit Inoperable Verify with the customer that they are using a good source of power outlet If this is not the issue ...

Страница 11: ... an acceptable leak rate is 10 If leak is not the issue contact Somnetics Customer Service P8 Battery Non functioning LED Light The customer should test the battery function by charging it for 8 hours and then using it until the battery dies This will tell us if it is a functional issue or just the LED light is no longer functioning Battery Usage The P8 battery should last approximately 14 hours w...

Страница 12: ...to Clinician from the previous screen The compliance file will automatically generate and save to a default location To find the file look at the line that says The file is located at This will be where the compliance file is saved d Attach the compliance report to an email to send 5 How to import a To view double click on the compliance file and it will load it into the Somnetics Sleep Apnea Ther...

Страница 13: ... top of the screen to get successively more detailed compliance information i Each of these screens have similar date range and export controls Device Not Recognized 1 First try disconnecting the USB and closing the software 2 Reconnect the USB and then reopen the software If this worked the CPAP serial number or green checkmark will show in the upper right Return to step 3 in the correct section ...

Отзывы: