Operation Manual for Hand-held Refractometer for Brix 
measurement

 

 

 
 
Product Layout

 

 
 
 
 
 
 

 
 
 

 
 
 
Specification: 

 

Model: 

RBX0010 

RBX0018 

RBX0032 

RBX0080 

RBX0090 

Measuring range: 

0 ~10% 

0 ~18% 

0 ~ 32% 

0 ~ 80% 

0 ~ 90% 

Display Resolution:  0.1% 

0.1% 

0.2% 

0.5% 

0.5% 

 
Temperature Compensation:  Automatic from 10 to 30degree Celsius except 
RBX0080 & RBX0090 
 
 

Calibration:

  

1. 

Calibration should be made in an ambient environment of 20degree Celsius.  

2. 

Open the Light Plate & place few drops of distilled water ½ fill the prism. 

3. 

Close the Light Plate.  Ensure the water is spread across the entire surface of 
the Prism.  There must not be any bubbles or dry spots on the Prism, which is 
cause by too little liquid use. 

4. 

Wait  for  about  30  seconds;  this  is  to  allow  the  temperature  compensation  to 
take place. 

5. 

Hold the meter in the direction of light source and look into the Eyepiece. 

6. 

Adjust the Focus Dial until image is clear and sharp.  Each person has his own 
focusing adjustment. 

7. 

Remove the rubber cap on the Calibration knob,  with the use  of the  provided 
trimming tool and adjust it until the bottom line of the blue section is at 0%. 

8. 

Calibration Completed. 

9. 

Wipe  dries  the  Light  Plate  and  Prism  with  soft  lint-free  clothes  before 
proceeding with measurement. 

 
 
 
 
 

Operation Manual for Hand-held Refractometer for Brix 
measurement

 

 
 
 

Product Layout

 

 
 
 
 
 
 

 
 
 
 

 
 
Specification: 

 

Model: 

RBX0010 

RBX0018 

RBX0032 

RBX0080 

RBX0090 

Measuring range: 

0 ~10% 

0 ~18% 

0 ~ 32% 

0 ~ 80% 

0 ~ 90% 

Display Resolution:  0.1% 

0.1% 

0.2% 

0.5% 

0.5% 

 
 
Temperature Compensation:  Automatic from 10 to 30degree Celsius except 
RBX0080 & RBX0090 

 
Calibration:

  

1. 

Calibration should be made in an ambient environment of 20degree Celsius. 

2. 

Open the Light Plate & place few drops of distilled water ½ fill the prism. 

3. 

Close  the  Light  Plate.    Ensure  the  water  is  spread  across  the  entire  surface  of 
the Prism.   There  must  not  be  any  bubbles or  dry spots on  the Prism,  which is 
caused by too little liquid use. 

4. 

Wait for about 30 seconds; this is to allow the temperature compensation to take 
place. 

5. 

Hold the meter in the direction of light source and look into the Eyepiece. 

6. 

Adjust the Focus Dial until image is clear and sharp.  Each person has his own 
focusing adjustment. 

7. 

Remove  the  rubber  cap  on  the  Calibration  knob,  with  the  use  of  the  provided 
trimming tool and adjust it until the bottom line of the blue section is at 0%. 

8. 

Calibration Completed 

9. 

Wipe dries the Light Plate and Prism with soft lint-free clothes before proceeding 
with measurement. 

 

Eyepiece 

Calibration Knob 

Focus Dial 

Light Plate 

Prism 

Eyepiece 

Calibration Knob 

Focus Dial 

Light Plate 

Prism 

Making Measurement: 

1. 

Open the light plate & place 2 to 3 drops of the test liquid on the prism. 

2. 

Close the Light Plate.  Ensure the water is spread across the entire surface of 
the Prism.  There must not be any bubbles or dry spots. 

3. 

Wait  for  about  30  seconds;  this  is  to  allow  the  temperature  compensation  to 
take place. (ATC models only) 

4. 

Hold the meter in the direction of light source and look into the Eyepiece. 

5. 

Select the scale, which is the intended unit of measurement. 

6. 

Take  the  reading  from  the  scale  on  the  line  where  the  blue  and  white  area 
meets. 

7. 

Wipe  dry  the  Light  Plate  and  Prism  with  a  soft  lint-free  clothe  after  each  and 
every test and before storing away. 

8. 

It is important to maintain the Light Plate and Prism dust and stain free in order 
to have good repeatability measurement. 

 

Precautions in Usage: 

1. 

Clean  the  instrument  between  each  measurement  and  before  storage  with  a 
soft cloth.  Residue of samples stained on the prism could damage the coating 
on the prism. 

2. 

Do  not  measure  with  abrasive  or  corrosive  chemicals.    This  will  damage  the 
prism’s coating. 

3. 

Do  not  immerse  the  instrument  in  water.    If  the  view  inside  the  Eyepiece 
becomes  foggy,  water  has  entered  into  the  body.    Call  a  qualified  service 
technician or contact your dealer. 

4. 

Do not clean the  prism  with thinner  as this will remove the special  coating on 
the prism and damage the unit. 

5. 

Do not store the instrument in a damp environment.  Store only in a cool  and 
dry environment. 

6. 

This is an optical instrument and  requires careful handling and storage.  With 
care, this instrument will provide years of reliable service. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

 

 

Making Measurement: 

1. 

Open the light plate & place 2 to 3 drops of the test liquid on the prism. 

2. 

Close  the  Light  Plate.    Ensure  the  water  is  spread  across  the  entire  surface  of 
the Prism.  There must not be any bubbles or dry spots. 

3. 

Wait for about 30 seconds; this is to allow the temperature compensation to take 
place. (ATC models only) 

4. 

Hold the meter in the direction of light source and look into the Eyepiece. 

5. 

Select the scale, which is the intended unit of measurement. 

6. 

Take  the  reading  from  the  scale  on  the  line  where  the  blue  and  white  area 
meets. 

7. 

Wipe  dry  the  Light  Plate  and  Prism  with  a  soft  lint-free  clothe  after  each  and 
every test and before storing away. 

8. 

It is important to maintain the Light Plate and Prism dust and stain free in order 
to have good repeatability measurement. 

 

Precautions in Usage: 

1. 

Clean the instrument between each measurement and before storage with a soft 
cloth.  Residue of samples stained on the prism could damage the coating on the 
prism. 

2. 

Do  not  measure  with  abrasive  or  corrosive  chemicals.    This  will  damage  the 
prism’s coating. 

3. 

Do  not  immerse  the  instrument  in  water.    If  the  view  inside  the  Eyepiece 
becomes  foggy,  water  has  entered  into  the  body.    Call  a  qualified  service 
technician or contact your dealer. 

4. 

Do not clean the prism with thinner as this will remove the special coating on the 
prism and damage the unit. 

5. 

Do not store the instrument in a damp environment.  Store only in a cool and dry 
environment. 

6. 

This  is  an  optical  instrument  and  requires  careful  handling  and  storage.    With 
care, this instrument will provide years of reliable service. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 