RECALL MINIMUM/MAXIMUM

1.  Press MIN/MAX to view the minimum and maximum

  temperatures recorded since turning the unit on or

  since clearing min/max.

2.  Press the MIN/MAX button once to display the

  minimum temperature. The lower portion of the

  display shows 

MIN

 indicating that this is the

  minimum reading. While in the MIN/MAX mode,

  select the desired format with the °F/°C or

  .0/.00/.000/.0000 buttons.

3.  Press the MIN/MAX button again to display the

  Maximum temperature. The lower portion of the

  display shows 

MAX

 indicating that this is the

  maximum reading. A third press will return the 

  display to the current reading.

4.  MIN/MAX may be reviewed at any time.

5.  To clear the MIN/MAX, place the unit in normal mode

  (not reviewing the MIN/MAX) and press MIN/MAX

  and CLEAR simultaneously. You may also clear 

  MIN/MAX by turning the unit off.

HOLD FUNCTION

1.  Press the HOLD button once to 

freeze

 the display

  at the current temperature reading. 

HOLD

 appears

  on the lower portion of the display indicating that the

  unit is in hold mode.

2.  While in the HOLD mode, select the desired format

  with the °F/°C or .0/.00/.000 buttons.

3.  Press the HOLD button a second time to return to

  the current temperature reading. 

HOLD

 will no

  longer appear on the display.

POWER

Do not turn the unit on and off rapidly. It may lock the 

display. When turning the unit on/off the microprocessor 

may become locked, inoperable, or the display may 

read 

888888.

 If this occurs, reset the thermometer by 

removing the battery, waiting 15 seconds, and replacing 

the battery.

OPERATION

1.  Plug the probe into the receptacle located on the right

  side of the unit.

2.  Press the ON/OFF button to turn the unit on.

3.  Press the °F/°C button to change the display 

  between Fahrenheit and Celsius.

4.  Press the .0/.00/.000/.0000 button to change the

  display to read the desired resolution.

5.  Use the probe to monitor temperatures in air/gas,

  liquids, and semi-solids. Place the stainless-steel 

  portion of probe in contact with the material to be

  measured. In most instances the depth of the probe

  needs to be approximately 

½

 inch. 

6.  Press the ON/OFF button to turn the unit off when the

  thermometer is not in use to prolong battery life.

HISTORY MODE

History provides an effortless method to observe when a 

temperature change takes place.  It continuously displays 

for the past 24 hours starting with 1 hour ago. If left on for 

more than 24 hours, it displays only the most recent 24 

hours. History may be reviewed at any time.

1.  After 1 hour, press the HISTORY button once to 

  show the current or first hour reading. The display

  will show 

1

 on the far right to indicate this is the first

  reading. After approximately two seconds, the display

  will show the temperature and 

min

 which indicates

  this is the minimum reading for the first hour.  After

  approximately two more seconds, the display will

  again show 

1

. The display continues to alternate

  between these two displays. While in the history

  mode, the °F/°C or .0/.00/.000 buttons may be used

  to select the desired format.

2.  The second press of the HISTORY button shows the

  maximum temperature for the current or first hour.

  After approximately two more seconds, the display

  will show 

1

.

3.  With each press of the HISTORY button, the unit 

  will scroll through all 48 minimum and maximum

  readings.

4.  To exit the history mode, press the CLEAR button 

  or MIN/MAX button.  As long as the unit is in the

  history mode, pressing the CLEAR button alone will

  not clear the history.

5.  To clear history, first exit the history mode and then

  press HISTORY and CLEAR buttons simultaneously.

  Turning the unit off does not clear the history.

SPECIFICATIONS

Range:

 

-328.00 to 932.000°F  

 

-200.00 to 500.000°C

Resolution:

 

0.001°, 0.01° or 0.1° 

(user selectable)

 

from -99 to 932°F/-99 to 500°C

 

0.01° or 0.1°

 

from -328 to -100°F/-200 to -100°C

 

0.0001° from -9.9999 to 99.9999 °F/°C

Sampling Rate:

  approximately 2 times per second

Memories:

 

48 Memories are stored in the History

 

Mode (min/max for the past 24 hours)

 

2 Memories are stored in the Min/Max

 

Mode (min/max for the current hour)

Probes Supplied:

 

Model 6412

 supplied with a fast response, handle-probe.  

The stainless steel portion of the probe is capable of 

measuring temperatures through the entire range (-328 to 

932°F/-200 to 500°C)

Model 6413

 supplied with a fast response, bullet-probe 

with PTFE cable.  The sensor 

and 

cable are capable of 

being subjected to temperatures of -148 to 392°F/-100 

to 200°C

BUTTON DESCRIPTION

  ON/OFF

 

Turns unit on and off.

 

°F/°C:

 

Selects °Fahrenheit/°Celsius

 .0/.00/.000/.0000

  Selects decimal point on display

  HISTORY

 

Displays history memories

 HOLD/CLEAR

  Press to hold reading also used in

 

 

conjunction with HISTORY or 

 

 

MIN/MAX buttons to exit the mode or

 

 

clear readings.

  MIN/MAX

 

Displays the highest and lowest

 

 

temperatures recorded for the current

 

 

hour.

DISPLAY MESSAGES

▲ 

Appearing indicates temperature is rising. 

▼ 

Appearing indicates temperature is falling.

LLL

” 

Displayed when the temperature being 

   

measured is below the range of the unit or when

   

there is an open probe or no probe.

HHH

   Displayed when the temperature being

   

measured is above the range of the unit or

   

when there is a shorted probe.

BAT

” 

Displayed when the battery is low and needs to

   

be replaced. See Battery section for battery

   

replacement instructions.

RECEPTACLES

The receptacle on the right side of the unit is for the 

probe. The upper left side receptacle is for an accessory 

AC Adaptor.  The lower left side receptacle is for the 

accessory Data Acquisition System data cable.

DATA OUTPUT

In order to use the data output, the 

Accessory Data 

Cable Cat No. 4099

 is required. See Accessories section 

for ordering information.

With the accessory data cable attached, to toggle the 

RS232 output on/off, press and hold the °F/°C and 

.0/.00/.000/.0000 simultaneously (

2320

 indicates 

RS232 output is 

on

232F

 indicates RS232 output 

is 

off

).

BATTERY

If 

BAT

 appears on the display, it indicates the 

batteries are low and need to be replaced. To replace 

the battery, slide the battery cover located on the back 

of the unit away from the unit. Remove the old battery 

and replace it with a new 9-Volt alkaline battery. Use an 

alkaline battery, NOT a regular or heavy duty battery. 

Properly connect the battery. Replace the battery cover. 

Incorrectly installed batteries may damage electronics.

ALL OPERATIONAL DIFFICULTIES

If this thermometer does not function property for any 

reason please replace the battery with a new 9-Volt 

alkaline battery Battery section, above). Low battery 

power can occasionally cause any number of 

apparent

 

operational difficulties. Replacing the battery with a new 

fresh battery will solve most difficulties.