Page 37
15. Initialising the WP-82Y
If the calibration settings of the
WP-82Y
exceed the allowable limits, the
unit may need to be initialised to factory default values. This action may
be required if the sensor is replaced.
To initialise the
WP-82Y
…
1. Switch the
WP-82Y
OFF.
2. Press
and HOLD
the
key while switching the
WP-82Y
back on.
3.
The following messages should be displayed…
Initialized
MUST ReCalibrate
then :
WP82Y V2.0 S1234
Oxygen Temp.
(The “s” after
WP-82Y
is shown when the RS232 serial port option is
fitted)
4. The meter then displays Dissolved Oxygen and Temperature. Note
that the decimal poi
nts have been replaced with a “
“, to indicate that
the unit requires re-calibration.
5. When the
WP-82Y
is initialised:
(a) The manual salinity correction value is re-set 36.0ppK. See
section 6 if you wish to change this value.
(b) The Altitude or Pressure correction system is switched off, and
the instrument assumes sea level (0m) and 1013 HPa.
16. Instrument firmware version number.
If you need to phone or fax TPS for any further technical assistance, the
version number of your
WP-82Y
firmware may of benefit to us. The
version number is displayed by the
WP-82Y
at turn-on.