Page 30
11.6 Data Format
Data is returned to the RS232 Port by the
WP-82Y
in the following
format. Please note that a “
” shown anywhere in this section denotes
one space.
LLLL
DDDDDDddd
SSSSSSppK
TTTTTToC
AAAAAAaaa
dd/mm/yy
hh:mm:ss
where…
LLLL
is the Log Number. Maximum 4 characters, right justified.
The
WP-82Y
sends Zero for instant readings (section 11.2)
DDDDDD
is the Dissolved Oxygen Data. Maximum 6 characters,
right justified.
ddd
is the unit description, either
“
ppM
”, “
%S
”, or “
%G
”
SSSSSS
is the Salinity correction value. Maximum 6 characters,
right justified.
ppK
is the salinity correction value unit description.
TTTTTT
is the Temperature Data. Maximum 6 characters, right
justified.
oC
is the Temperature unit description
AAAAAA
is the Altitude
OR
Pressure correction value. Maximum 6
characters, right justified.
aaa
is the unit description, either “m
” for Altitude or “HPa” or
Atmospheric Pressure.
dd/mm/yy
is the date, month and year data.
hh:mm:ss
is the hours, minutes and seconds data.
Notes
1. When requested by a PC with the ?D or ?R commands (section
11.5), the data is terminated with a carriage return.
2. When the data is sent by the
WP-82Y
using the Print function
(section 9.4) or the Instant Send function (section 11.2), the data
ends with a carriage return and a line feed.