11. Troubleshooting
11.1 General Error Messages
Error
Message
Possible Causes
Remedy
Factory
Cal. Fail
See
Handbook
The EEPROM chip which
contains the factory
calibration information has
failed.
The unit must be returned to
TPS for service.
Memory
Failed
Calib.
Lost
Memory
Reset !
You MUST
Re-Cal.
User calibration settings
have been lost or corrupted.
Re-calibrate the instrument.
Both a Zero and an Air
calibration are required for
Dissolved Oxygen (section
3), a 2 point calibration for
pH (section 4) and a 1 point
calibration for temperature
(section 5).
Meter displays the
word
OFF
, and
switches off.
Battery is below 6.00 volts.
Replace the battery.
Meter will not turn
on.
Battery is exhausted.
Replace the battery.
Flashing
symbol.
Battery is below 7.50 volts.
Replace the battery soon.
Note that the unit will switch
itself off when the battery
falls below 6.00 volts.