background image

 

 
Safety Considerations 
Warning: Please follow the manufacturer’s test procedure whenever possible. Make 
sure the pressure to be measured is not outside the range of the TPI 610/615/620/625 
meter 

 

General Guidelines 
Always 

·

 

Test the pressure meter before use to ensure it is operation properly 

·

 

Ensure that the connection hose you are using is free from kinks or splits 

·

 

Double check all connections before testing 

Never 

·

 

Connect the Instrument to an unknown pressure source if the source is twice the 
working pressure of the instrument the sensor will be damaged and may cause 
injury to the user 

·

 

Tamper with the instrument or attempt to modify it in any way. Otherwise 
accident and injury may occur and the warranty becomes void. 

 
page 4 
 

www.

.com

[email protected]

1.800.561.8187

Содержание 600 Series

Страница 1: ...I 610 Single Input 4 Function single unit of pressure TPI 615 Single Input 7 Function 7 units of pressure TPI 620 Dual Input differential 4 Function single unit of pressure TPI 625 Dual Input differential 7 Function 7 units of pressure All these models measure over the range of 200 0 to 200 0 mbar Each unit has increased resolution of 0 01 mbar and automatically auto ranges at 19 00 mbar to 0 1 mb...

Страница 2: ...rly Ensure that the connection hose you are using is free from kinks or splits Double check all connections before testing Never Connect the Instrument to an unknown pressure source if the source is twice the working pressure of the instrument the sensor will be damaged and may cause injury to the user Tamper with the instrument or attempt to modify it in any way Otherwise accident and injury may ...

Страница 3: ...ssure readings MAX Enables the user to recall display both the MIN maximum and minimum pressure recorded since the function was selected Press button once to actuate twice to read minimum and three times to read maximum To cancel the function press Zero TPI 615 625 only HOLD Maintains the reading on the LCD display TPI 615 625 only UNITSSelects 7 units of pressure kPa PSI inHg mmHg inH20 mbar TPI ...

Страница 4: ... Function Provide immediate push button zeroing for Speed and accuracy Record Max Min Enables long term testing of pressure circuit to check for leaks TPI 615 625 only Hold Enables the user to freeze the display reading when working in cramped working conditions when the display is not visible TPI 615 625 only Units Change the display value in any one of the following engineering units kPs PSI inH...

Страница 5: ... 2 2 036 3 937 x 10 2 2 89567 x 10 3 2 953 x 10 4 inH20 4 C 406 8 401 48 13 60 1 27 68 0 535 4 3 937 x 10 2 4 014 x 10 3 Lb in2 PSI 14 69559 5 14 504 0 4912 3 612 7 x 10 2 1 1 934 x 10 2 1 42232 3 x 10 3 1 4503 x 10 4 mmHg torr 760 750 06 25 400 1 868 51 715 1 7 35433 x 10 2 7 502 x 10 3 mmH2 0 10 33272 x 10 3 10 197 6 x 10 3 3 4544 x 10 2 25 40 0 703 07 13 59 9 1 0 101955 6 Pascals 1 01325 x 10 5...

Страница 6: ... the backlight in poor light conditions press the ON key at any time If the reading is erratic or changing activate the smoothing function by pressing the key To freeze a reading in awkward or cramped working spaces press the HOLD key TPI 615 625 only To record maximum minimum pressures when establishing pressure decay or carrying out gas line soundness testing press the MAX MIN key TPI 615 625 on...

Страница 7: ...o Range 200 0 mbar to 200 0 mbar Accuracy 0 5 of reading 5 Digits Over Pressure 500 mbar Power Supply 9V Relative Humidity 0 to 80 0 C to 40 C Operating Temp 0 C to 40 C Storage Temp 10 C to 50 C Auto Power Off 20 Minutes Input Type Single Single Differential Differential Trim Mode Yes Yes Yes Yes Zero Adjust Yes Yes Yes Yes Record Min Max No Yes No Yes Units 1 7 1 7 Data No Yes No Yes In units th...

Страница 8: ...nstrument 3 Remove the back cover 4 Remove old battery 5 Replace with new battery 9V NEDA 1604 or equivalent 6 Reassemble the instrument in reverse order from above being sure not to trap any wires between the two helves of the case 7 Re fit the six cross head screws being sure not to over tighten Cleaning Use a mild detergent and a slightly damp cloth to clean the surfaces of the instrument page ...

Страница 9: ...This is to certify that model 610 615 620 625 conform to the protection requirements of the council directive 89 336 EEC in the approximation of laws of the member states relating to Electromagnetic compatibility be application of the following standards EN50081 1 1992 Emissions Standard EN50082 1 1992 Immunity Standard To ensure conformity with these standards this instrument must be operated in ...

Страница 10: ... to the product which are the result of abuse improper use or maintenance are not covered Any other expense consequential damages incidental expenses including damages to property is not covered Transportation expenses to the company are not covered To obtain warranty performance include with the product your name address phone number written description of the problem and proof of purchase date C...

Отзывы: