9
8
e. OHM (Resistance, )
Range
Res.
Accuracy
Overload
Protection
400
Ω
0.1
Ω
±1% of reading, ±10 digits
600V DC or
AC Peak
f. Continuity Buzzer
Test Voltage
Threshold
Over Load Protection
3V
< 50
Ω
(100 digits)
600 V DC or Peak AC
h. General Specifications
Max. Volt. between
600V
any Input and Ground
Display Type
4,000 Count, 4 times per
second update
Operating Temp.
-10° to 45°C (14° to 113°F)
Storage Temp.
-20° to 55°C (-4° to 131°F)
Relative Humidity
0% to 80% (0° - 35°C/32° - 95°F)
0% to 70% (35° - 50°C/95° - 122°F)
Power Supply
9 Volt Battery
Battery Life
80 hrs. Alkaline
Size (H x L x W)
40mm x 190mm x 65mm
(1.5in x 7.25in x 2.5in)
Weight
315g (11.1oz)
3. Specifications
IEC 1010 Over Voltage:
CAT II - 1000V
CAT III - 600V
Pollution Degree 2
a. DCV
Range
Resolution
Accuracy
Impedance
40V
0.01V
±0.75% of reading,
10M
Ω
400V
0.1V
±3 digits
600V
1V
b. ACV
Range
Resolution
Accuracy
Impedance
400V
0.1V
±1.2% of reading,
10M
Ω
750V
1V
±3 digits
c. ACA
Range
Res.
Accuracy
Freq. Response
400A
0.1A
±1.5% of reading,
45Hz to 450Hz
700A
1A
±4 digits
UL 3111-1
*Warning:
Test Leads. Use only correct type and
overvoltage category rating.
!