1010a
Rev. 3.x & Above Meters
Indoor Air Quality Meter
The Value Leader
TM
www.
GlobalTestSupply
.com
Find Quality Products Online at:
[email protected]
Страница 1: ...1010a Rev 3 x Above Meters Indoor Air Quality Meter The Value LeaderTM www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...
Страница 2: ...des 4 5 Activating the Backlight 4 6 Printing Measurements 4 7 Setting Date and Time 4 8 Setting CO Alarm Level 4 9 Ambient Air CO2 Reset 5 Calculating Outside Air 5 1 Outside Air Temperature 5 2 Outside Air Carbon Dioxide 6 Data Logging 7 Retrieving Logged Data 8 Technical Specifications 9 Maintenance Troubleshooting 9 1 Battery Replacement 9 2 Service 10 Technical Information www GlobalTestSuppl...
Страница 3: ... installed in the unit Therefore the time the unit is off and not being used must be taken into account when determining sensor life The sensors in your analyzer are warranted for two years This warranty does not cover sensors damaged through misuse of the meter You should keep battery power applied to your sensors at all times The following guidelines will help prevent damage to your sensors Alwa...
Страница 4: ...ual Your TPI 1010a Indoor Air Quality meter has the following optional accessories available Power adapter Battery eliminator A804 Infrared Printer A740 Magnetic strap kit to hang meter A127 Boot hook to hang meter A103 Replacement Software only A802 Replacement USB cable only A801 Denotes part number A804 Battery Eliminator Specifications Input 100 240VAC 47 63Hz Output 5VDC 2A Center pin positiv...
Страница 5: ...nnection of USB cable for PC communication Power Adapter Socket Location to connect optional power adapter Infrared Window Used to print test results to optional infrared printer Humidity Temperature Sensor Location Display Keypad CO2 Sensor Location CO Sensor Location USB Port Power Adapter Protective Rubber Boot Infrared Window for Printing Test Results Requires optional A740 printer www GlobalT...
Страница 6: ...d change parameters Enter key Used to activate outside air mode Also used to accept parameter changes Mode key Used to activate and deactivate data logging mode Press and hold to activate or deactivate logging Backlight key Used to turn the display backlight on and off T C key Used to switch the bottom display between temperature humidity and time display modes Also used to perform ambient air CO2...
Страница 7: ...rument for hands free operation Can be used with A127 magnetic strap kit or A103 hook Tilt Stand Use to stand instrument on a table Batter Cover Loosen this screw and open cover to access batteries Screw www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...
Страница 8: ... 30ppm or more CO is measured the 1010a will beep and the display will flash red as a warning This alarm point is adjustable Additional functions can be activated during testing Please see next sections Carbon dioxide CO2 Carbon monoxide CO Temperature Humidity Battery indicator Press the key and the 1010a will begin a 30 second countdown During this time the 1010a performs a self diagnostic and s...
Страница 9: ... cycles between temperature and clock display To set the clock please refer to section 4 6 Pressing the key turns the display backlight on and off Power consumption is much higher when the backlight is activated The backlight will automatically turn off 30 seconds after it is activated To conserve bat tery life the backlight should only be used when necessary Align the infrared window of the A740 ...
Страница 10: ...s set press 5 Using the Arrow keys set the hour Once the hour is set press 6 Using the Arrow keys set the minutes Once minutes is set press The 1010a will return to normal operation key then press and hold down the key until two beeps are heard and the time date screen is displayed Year Month Day Hour Minutes 24 hour clock HH MM 4 7 Setting Date and Time www GlobalTestSupply com Find Quality Produ...
Страница 11: ...eys until the CO alarm screen is displayed 4 9 Ambient Air CO2 Reset The CO2 sensor can be reset at ambient level as needed This can be performed if the CO2 reading seems to be too high or low or just to ensure your 1010A is operating properly Note This does not take the place of a full factory calibration which is recommended once per year 1 Take the 1010A outside in ambient air Allow the CO2 rea...
Страница 12: ... air temperature t 1 is displayed 2 Place the sensor of the 1010a perpendicular to the Supply air flow When the reading stabilizes press the key and capture the supply air temperature t 2 is displayed 3 Place the sensor of the 1010a perpendicular to the Outside air flow When the reading stabilizes press the key and capture the outside air temperature t 3 is displayed 4 Press the key t will be disp...
Страница 13: ...supply air temperature C 2 is displayed 3 Place the sensor of the 1010a perpendicular to the Outside air flow When the reading stabilizes press the key and capture the outside air temperature C 3 is displayed 4 Press the key C will be displayed and the calculated percent outside air will be displayed at the top of the display 5 If you want to perform this measurement again press the key once and r...
Страница 14: ...1 sec to 59 min 59 sec In the minutes mode the time is adjustable from 1 min to 2 Once the time mode is set press the keys select the digit and the keys increase or decrease the value key and time digits will flash Using the Arrow keys set the interval to the desired level The 3 Once the seconds have been entered press the key to move to min hr 5 Once the logging interval is set press the keys sel...
Страница 15: ... open it Connect the 1010a to the PC using the USB cable Click on the connect to PC button in the software window The software can be used to retrieve logged data and for real time monitoring COM port setting is Baud rate 19200 Parity None Data Bits 8 Stop Bit 1 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...
Страница 16: ...3hr 59min interval Display Type 3 line with annunciators and backlight CO Alarm Audible and visual selectable level factory default 30ppm Operating Temperature Humidity CO and CO2 14 F to 122 F 10 C to 50 C All Other Function 4 F to 140 F 20 C to 60 C Humidity 15 95 non condensing Storage Temperature 4 F to 140 F 20 C to 60 C Power supply 3 AA batteries 1 5V each x 3 4 5V total Battery Life 40 hou...
Страница 17: ...calibration or replacement Corrective Action Replace batteries Send unit to TPI for factory calibration and maintenance 9 1 Battery Replacement When the batteries become low they will require replacement 1 Turn the meter over so the back is facing you 2 Loosen the screw holding down the battery cover located under the tilt stand 3 Lift the tilt stand up and remove the battery cover 4 Replace the b...
Страница 18: ...y reduce the amount of outdoor air that enters a home pollutants can build up even in homes that are normally considered leaky Parameter IDPH1 ASHRAE2 OSHA PEL3 ACGIH TLV4 Humidity 20 60 30 60 N A N A Temperature 68 75 winter 73 79 summer 68 75 winter 73 79 summer N A N A Carbon Dioxide 1 000ppm 800ppm preferred 1 000ppm 5 000ppm 5 000ppm Carbon Monoxide 9ppm 9ppm 50ppm 25ppm N A Not Applicable or...
Страница 19: ...ection 9 2 for instruc tions on returning your meter CO Sensor Caibration Due When the last calibration date of the CO sensor is one year or older the 1010a will display a reminder that calibration is due before returning to normal operation Please return your instrument to TPI for calibration See section 9 2 for instructions on returning your meter Temperature and or Humidity Sensor Error If the ...