Instruction Manual - TORBAL DRX 5 Series
5. If the stability indicator appears and the scale does not sound a beep nor the “2% DEV” or “3% DEV" command
appears on the display, a deviation of greater than 2% was not found.
Note: If a deviation exists between the stored average piece weight and the average piece weight of the counted pills,
update the average piece weight before continuing counting in order to avoid possible counting errors. For details on
maintaining and modifying the average piece weight turn to chapter 16 of this instruction manual.
CHAPTER 20.
ADVANCED PILL COUNTING ACCURACY (APA)
When enabled APA can adjust average piece weight variations that may exist between a currently counted drug, and the
average piece weight of that drug that is stored in the scale’s database. To use the APA feature follow the steps below.
1. Once command “FILL” is displayed slowly begin to count the pills. In order to activate APA stop the count
between 10 and 15 pieces.
Note: When stopping outside of the 10 to 15 range the feature will not work.
2. When you reach a count of 10 to 15 pieces stop and wait for the scale to stabilize.