4. Operation
Page 43
Status: 6.9.18
4.4
Power Up of FemtoFiber
®
pro Systems
For safe operation of the FemtoFiber
®
pro Laser System please follow the steps below.
1.
Make sure all connections according to section 4.3 are installed.
2.
Switch ON the Mains Switch on the rear panel of the FemtoFiber
®
pro Control Unit(s) to set the sys-
tem in standby mode.
3.
Switch ON the FemtoFiber
®
pro Control Unit(s) by the Key Switch at the front panel and wait until
the boot sequence is finished. This is indicated by an acoustic signal and the green Power Status
LED lighting continuously.
4.
Manual Operation:
To activate/deactivate the laser(s) press the yellow Laser Emission Start/Stop button on the
FemtoFiber
®
pro Control Unit(s) front panel.
Computer Controlled Operation:
Start the FemtoFiber
®
pro GUI(s) and connect each FemtoFiber
®
pro Control Unit as described in
section 5.4. Click the
ON
or
OFF
button on the
Laser Control
Tab of the FemtoFiber
®
pro GUI to
activate/deactivate the laser controlled by the respective GUI.
5.
The white Laser Radiation Emission Warning LED (Figure 30: 3) lights up continuously and indicates
laser emission.
Open the desired shutter.
4.5
Power Down
1.
To switch OFF the FemtoFiber
®
pro Laser System, deactivate the laser(s) at the respective yellow
Laser Emission Start/Stop button on the FemtoFiber
®
pro Control Unit(s) front panel.
2.
Switch OFF the FemtoFiber
®
pro Control Unit(s) by the Key Switch on the front panel.
DANGER !
Do not look into the exiting beam of the FemtoFiber
®
pro Laser Head as the output can
exceed the limits for class 1 specified by the US laws 21 CFR 1040.10 and 2 CFR 1040.11 and
the Laser Safety Standard EN 60825-1:2014. Take precautions to eliminate exposure to a
direct or reflected beam.
NOTE !
The FemtoFiber
®
pro Control Unit is equipped with an universal mains input which sets the
FemtoFiber
®
pro Control Unit to your local mains supply voltage.
NOTE !
The operation of systems with VAR option is described in section 6.2 in the appendix.
CAUTION !
With a
Multi-Arm Laser System
, for safety reasons and to guarantee optimum system per-
formance, before powering-up always check whether the FC/APC fiber connections are
installed properly.
NOTE !
The operation of systems with VAR option is described in section 6.2 in the appendix.
Содержание FemtoFiber pro IR
Страница 2: ......
Страница 4: ......
Страница 8: ...FemtoFiber pro Laser System Status 6 9 18...
Страница 50: ...FemtoFiber pro Laser System Page 46 Status 6 9 18 Figure 35 7 Click Finish to complete the installation...
Страница 98: ...FemtoFiber pro Laser System Page 94 Status 6 9 18 6 10 Declaration of CE Conformity FemtoFiber pro...
Страница 99: ...6 Appendix Page 95 Status 6 9 18 6 11 Declaration of CE Conformity FemtoFiber pro SCYb...
Страница 107: ......
Страница 108: ......