OPTIONAL OPERATIONS
MANUAL MODE IN TONO/PACHO
SETTING THE MANUAL MODE
1
Check that the MEASUREMENT screen is on. If the
button is "M," the mode is
Manual mode.
2
If "A" (Auto mode) is displayed, tap it and change to "M".
SETTING THE MEASURING RANGE
In this instrument, the measuring range can be switched in 2 steps between "1-30" and "1-60."
Normally, "1-30" is used, but if the patient's ocular pressure is high, switch it to "1-60." The default
setting is "1-30" upon power on.
1
Check the measurement screen.
2
Tap the
button of the control panel and set the measuring range.
NOTE
• Adjust the height of the instrument table so that the patient can sit com-
fortably. Otherwise, correct measurement values may not be obtained.
• Do not perform measurement if the patient holds their breath or is ner-
vous. Otherwise, correct measurement values may not be obtained.
• If the machine is moved before measurement values are displayed, it may
cause incorrect measurement results.
• The beep function for urging cautions is provided so that a finger or a
hand is not pinched between the reverse side of forehead rest, a measur-
ing head and an ocular pressure measurement window.
Auto/Manual
Patient ID
Patient ID
1
1
TOPCON̲TAROU
TOPCON̲TAROU
Focus and
Touch a measure point
T/P
MANUAL
MODE
30/60
Patient ID
Patient ID
1
1
TOPCON̲TAROU
TOPCON̲TAROU
Focus and
Touch a measure point
T/P
MANUAL