3.11 Average Measurement
Set whether average measurement should be performed or not.
Normal Meas: Average measurement is not done.
Average Meas: Average measurement is done.
In average measurement, the average value of the measured values obtained by measuring two or
more times is regarded as the measured value to perform measurement with low luminance more
accurately. The more times averaging is carried out, the longer the measuring time is.
Measuring time
= Integral time × (Times of av1) + Integral time + Shutter operating time + Processing time
Setting whether average measurement should be done or not
1.
Shift to the function mode. Access [MEASURE]
from the function menu. Then, access the
[AverageMeas ON/OFF] screen.
☞
“3.1.2 Entering/Returning from the Function Mode”.
2.
To change the setting, press the [CHANGE] switch. Press the [ROTATION] switch to
change “Normal Measure” and “Average Measure” to each other.
“Normal Meas” is changed to “Average Meas” and vice versa.