37
BASIC OPERATIONS
PREPARATION OF PATIENT
1
Check the capture screen.
2
Make sure that Photographing window is at the height of the eye height mark. Move the photo-
graphing head until photographing window is at the height of the eye height mark, if photograph-
ing window is above the eye height mark press the lower side of the control panel display, or if it
is beneath the eye height mark press the upper side of the control panel display.
3
Have the patient sit in front of the instrument.
4
Adjust the adjustable instrument table or the chair height for the patient to put his/her chin on the
chinrest comfortably.
5
Place the patient's chin on the chinrest and check that his/her forehead is touching to the fore-
head rest.
CAUTION
• To avoid electric shock, do not touch the external connection ter-
minal and the patient at the same time.
• To avoid injury, do not insert fingers under the chinrest.
To avoid injury when moving the chinrest down, be careful not to
catch the patient's finger. Tell this to the patient, too.
• To avoid injury when operating the machine (for measurement
and control panel operation), be careful about the cover not to
catch fingers of the patient. Tell this to the patient, too.
NOTE
• Adjust the height of the adjustable instrument table so that the patient can
sit on the chair comfortably. Otherwise, correct measurement values may
not be obtained.
• When operating the instrument (for measurement and control panel oper-
ation), be careful that the instrument does not touch the patient's lip or
nose. If touched, clean the instrument following "CLEANING THE FORE-
HEAD REST AND CHIN REST" on page 83.
• If no patient ID is registered, a "patient No." is assigned and displayed
automatically in order of examination.
Eye height mark of
photographing window
Eye height mark
Содержание SP-1P
Страница 1: ...USER MANUAL SPECULAR MICROSCOPE SP 1P ...
Страница 2: ......
Страница 18: ...16 COMPONENTS 5 Tap the to return to the capture screen check that patient ID is updated Next button 1081 ...
Страница 25: ...23 COMPONENTS SETUP SCREEN ...
Страница 90: ......
Страница 92: ...SPECULAR MICROSCOPE SP 1P 41904 97030 Printed in Japan 1308 100LW0 ...