17

ASSEMBLY PROCEDURE

* With the unit model, an accessory box is supplied to store the spare parts.

(11) Mounting the Hruby Lens and Hruby Lens Guide Plate (Option)

Mount the Hruby lens holder and the Hruby lens rail unit onto the chinrest assembly, using
four screws.

* Insert the Hruby lens and Hruby lens holder, which is located on the chinrest assembly.

Do not touch the lens surface.

* Place the guide plate in the hole of the main body's carrier arm.  The pointed end should face

the chinrest.

Hruby lens holder

Hruby lens rail

Hruby lens holder

Hruby lens

Содержание SL-3E

Страница 1: ...INSTRUCTION MANUAL SLIT LAMP SL 3E ...

Страница 2: ...tem to enhance natural color and resolution Robust construction for performance and durability This Instruction Manual covers an overview of the basic operation troubleshooting checking maintenance and cleaning of the SL 3E Slit Lamp To get the best results from this instrument read Displays for Safe Use and Safety Cautions Keep this Instruction Manual close at hand for future reference Warning To...

Страница 3: ...ns are appropriate 3 Do not store or transport the instrument on a slope or uneven surface or in an area where it is subject to vibrations or instability 4 Do not store the instrument where chemicals are stored or gas is generated 3 Usage period 8 years from delivery providing regular maintenance is performed according to the self cer tification TOPCON data ENVIRONMENTAL CONDITIONS FOR PACKAGING I...

Страница 4: ...N Ignoring or disregarding this display may lead to personal injury or physical damage Injury refers to cuts bruises sprains fractures burns electric shocks etc Physical damage refers to damage to buildings equipment or furniture ICON MEANING This indicates Prohibition Specific content is expressed with words or an icon either inserted in the icon itself or located next to the icon This indicates ...

Страница 5: ...ers and may affect other devices in the vicinity If you have discovered that turning on off the instrument affects other devices we recommend you change its position keep a proper distance from other devices or change the outlet Please consult the dealer from whom you purchased the instru ment for any additional questions TOPCON shall not take any responsibility for damage due to fire earthquakes ...

Страница 6: ...l appropriée afin d éviter le contact avec les yeux et le nez du patient CAUTION y To prevent electric shock switch off the power supply and remove the power cable before replacing the bulb Afin d éviter tout choc électrique coupez le contact et débranchez le câble d alimentation avant de remplacer l ampoule y To avoid burns do not touch the bulb immediately after switching the instrument off Afin...

Страница 7: ... TARGET 20 BASE OPERATION 21 OPERATION OF THE ILLUMINATION UNIT 22 FUNDUS OBSERVATION WITH THE HRUBY LENS OPTION 26 TROUBLESHOOTING 27 TROUBLESHOOTING GUIDE 27 SPECIFICATIONS 28 ELECTROMAGNETIC COMPATIBILITY 30 SYSTEM CLASSIFICATION 34 PURPOSE OF USE 34 OPERATION PRINCIPLE 34 SHAPE OF PLUG 35 SYMBOL 35 MAINTENANCE 36 DAILY CARE 36 REPLACING THE BULB 36 REPLACING THE FUSE 38 REPLACING THE CHINREST ...

Страница 8: ...st adjuster Hruby lens guide plate Option Illumination arm locking knob Click stop roller Angle scale Microscope arm locking knob Base relay cable Brightness adjustment knob Pilot lamp Power switch Lamphouse Aperture and slit length display window Filter selection lever Aperture and slit length control knob Fixation target Mirror Hruby lens Option Hruby lens holder Option Slit centering knob Illum...

Страница 9: ...s are the quantities Spare fixation target bulb 1 Focusing test rod 1 Spare illumination bulb 1 Spare chinrest paper pin 2 Spare fuse Taped onto the rear of the power supply 3 Brush 1 Dust cover 1 Instruction manual 1 Accessory box supplied with Unit Model Slit Lamp 1 16 eyepieces 2 Fixation target 1 Screwdriver small 1 SL 3E ...

Страница 10: ...9 CONFIGURATION Short mirror 1 Screwdriver large 1 Spare long mirror 1 Phillips screwdriver 1 Spare socket 1 Wrench not supplied with Unit Model Slit Lamp 1 ...

Страница 11: ...nation assembly 2 Chinrest assembly 3 Chinrest paper 4 Breath shield 5 Arm hole cover 6 Instrument type table w power supply 6 Unit type table w power supply 7 Rail cover 8 Base assembly 9 Microscope with 10 magnification eyepieces 10 Power cable ...

Страница 12: ...ly 1 2 Chinrest assembly 1 3 Chinrest paper 1 4 Breath shield 1 5 Arm hole cover 1 6 Instrument type table w power supply 1 6 Unit type table w power supply 1 7 Rail cover 2 8 Base assembly 1 9 Microscope with 10 magnification eyepieces 1 10 Power cable 1 ...

Страница 13: ... included wrench b To secure the unit model table to the ophthalmic unit Peel off the tape which secures the plastic washer to the mounting bracket s shaft Insert the shaft of the mounting bracket into the hole of the arm on the ophthalmic unit The plastic washer should be between the mounting bracket and the arm Note The power supply is attached to the left side of the unit top practitioner s vie...

Страница 14: ...mbly While making sure that the cord is not being pinched by the mounting plate re tighten the previously removed screws 3 Mounting the Base Assembly and Rail Covers After checking to see that the main unit moves smoothly on the rails remove the four screws which are attached to the rails with the screwdriver Place the rail cover to the rail and retighten the previously removed screws Chinrest cor...

Страница 15: ... secure the illu mination arm to the shaft of the base assembly 5 Mounting the Microscope The microscope body has a notched groove and tapped screw hole in the bottom of the cen tral barrel section While firmly holding the microscope from the top place the body assem bly on the microscope arm matching the notched groove with the stopper on the arm Support the microscope with one hand and tighten t...

Страница 16: ... the Illumination Arm Shipping Pad The pad is attached to protect the slit closure mechanism of the illumination arm assembly during shipping Remove the rubber band and gently pull the pad out 8 Connecting Cables a Connect the cable from the top of the chinrest to the lamphouse cover on illumination arm Peel off the tape which secures the lamphouse cover during shipping Breath shield attachment sc...

Страница 17: ... cables and re attach them to the table 9 Installing the Chinrest Paper Remove the two pins from the chinrest Place the pins through the holes in the chinrest paper Remove the wrapping from the chin rest paper Align the pins with the holes in the chinrest and secure the paper to the assembly 10 Spare Parts With the table model an accessory drawer is provided to store the spare parts Cable clip ...

Страница 18: ...y lens holder and the Hruby lens rail unit onto the chinrest assembly using four screws Insert the Hruby lens and Hruby lens holder which is located on the chinrest assembly Do not touch the lens surface Place the guide plate in the hole of the main body s carrier arm The pointed end should face the chinrest Hruby lens holder Hruby lens rail Hruby lens holder Hruby lens ...

Страница 19: ... on and observe that the illumination is passing through the opened slit controls Check to insure that the fixation device is illuminated Check to see that the slit width and length controls filter lever and magnification changer lever operate smoothly and properly 3 After the installation is completed turn the power supply off and cover the instrument with the dust cover ...

Страница 20: ...e Hruby lens guide plate Place the rod in the hole and turn it so that the flat black surface is perpendicular to the microscope s objective lens 2 Preparation of illumination unit Turn the power switch ON and place the brightness adjustment knob in an intermediate posi tion then set the slit width control knob so that slit is approximately 2 to 3mm wide 3 Diopter compensation To establish the cur...

Страница 21: ...so that the image is fused and a stereoscopic image results For operator s comfort since each eyepiece can be moved independently insure that both eye pieces are at the same height PATIENT POSITION AND FIXATION TARGET 1 Positioning patient s head Have the patient place his chin on the chinrest and forehead against the forehead rest Adjust the chinrest adjuster so that the patient s outer canthus i...

Страница 22: ...eeding 15D To replace the target loosen the locking screw and remove the target assembly Avoid over loosening the screw or it may drop Luminous fixation target BASE OPERATION 1 Horizontal gross adjustment To adjust the microscope s position horizontally move the base while keeping the joystick in the vertical position 2 Horizontal fine adjustment For fine adjustment such as alignment or focusing t...

Страница 23: ...erations described in 2 and 3 above while looking through the microscope OPERATION OF THE ILLUMINATION UNIT 1 Altering the slit size By operating the slit width control knob the slit width can be changed from 0 to 9mm at the 9mm size the slit becomes a circle The scale should be used simply as a guideline Your examination purpose will dictate the correct slit size Scale of slit width ...

Страница 24: ... slit image the slit length can be changed continuously from 1 to 8mm with this knob The values obtained are indicated through the display window 3 Rotating the slit image By moving the aperture and slit length control knob horizontally the slit image is rotated from the vertical through any oblique angle to the horizontal The angle of image rotation is indicated by the rotation angle scale Rotati...

Страница 25: ... particular benefit in indirect retro illu mination techniques To return the slit image to the center of the field of view tighten the slit centering knob firmly 5 Oblique illumination Oblique illumination is used for sectional or fundus examination by use of a contact lens etc i e gonioscopy As the inclination lever is released and the arm is pulled towards you the illumination unit inclines by 5...

Страница 26: ...0 Replacing the mirror 1 Set the angle between the microscope and illumination arms to exceed 30 2 Incline the illumination column at 10 or more 3 Remove the long mirror by holding on to the extended surface 4 Insert the beveled side of the short mirror 5 When removing the short mirror since it has no extension use an object with a sharp end to carefully push it out as shown below 7 Filter selecti...

Страница 27: ... shank in the groove on the guide plate 4 Center the illumination and microscope arms so that they face the patient s eye 5 Align lever B shown in the illustration below with the center of the microscope s field of view Then move the lever B backward and forward to position close to the patient s eye 6 Use the joystick to focus on the fundus The slit width and length will have to be adjusted to re...

Страница 28: ...et P 36 The bulb has burned out Replace the bulb P 36 Fuse has blown Replace the fuse P 38 Slit light is too dim The bulb is not correctly inserted Insert bulb correctly P 36 Filter lever is at ND position or at an intermediate position Set the filter lever to the correct position P 25 Fuse has blown Type of fuse used is incorrect Replace with the correct type as specified P 36 Slit width closes b...

Страница 29: ...ntinuous from 9mm to 0mm at 9mm slit becomes a circle Slit length Continuous from 8mm to 1mm OR Aperture settings at 9 8 5 3 2 1 and 0 2mm dia Slit angle 0 to 180 with horizontal scanning capability Slit inclination 5 10 15 and 20 steps Filters Blue red free 13 ND and heat absorbing Lamp 6V 20W Halogen lamp BASE Longitudinal movement 90mm Lateral movement 100mm Fine base movement with joystick 15m...

Страница 30: ...29 SPECIFICATIONS Subject to change in design and or specifications without advance notice DIMENSION WEIGHTS Table model 550mm 370mm Unit model 440mm 350mm Weight 21kg Table model 20kg Unit model ...

Страница 31: ...djacent or stacked use is necessary the EQUIPMENT or SYSTEM should be observed to verify normal operation in the configuration in which it will be used Guidance and manufacturer s declaration electromagnetic emissions The SL 3E is intended for use in the electromagnetic environment specified below The customer or the user of the SL 3E should assure that it is used in such an environment Emissions ...

Страница 32: ...erential mode 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and Voltage variations on power supply input lines IEC 61000 4 11 5 Ut 95 dip in Ut for 0 5 cycle 40 Ut 60 dip in Ut for 5 cycles 70 Ut 30 dip in Ut for 25 cycles 5 Ut 95 dip in Ut for 5 sec 5 Ut 95 dip in Ut for ...

Страница 33: ...mit ters as determined by an electromag netic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 NOTE 2 At 80 MHz and 800 MHz the higher frequency range applies These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection...

Страница 34: ...ration distance according to frequency of transmitter m 150kHz to 80MHz d 1 2 80MHz to 800MHz d 1 2 800MHz to 2 5GHz d 2 3 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended sepa ration distance d in meters m can be estimated using the equation applicable to the fre quency of the transmi...

Страница 35: ...h one comes into contact do not become conductive if the basic insulation fails The mode of operation continuous operation equipment Degree of protection against ingress of water IP 0 SL 3E is the ordinary instrument enclosed instrument without protection against ingress of water Methods of sterilization or disinfection recommended by the manufacture SL 3E does not have any part to be sterilized o...

Страница 36: ...Ecuador 110V 50Hz Type C E USA 120V 60Hz Type A Hospital Grade Canada 120V 60Hz Type A Hospital Grade Symbol IEC Publication Description Description French 60417 5032 Alternating Current Courant alternatif 60348 Attention consult accom panying documents Attention consulter les docu ments d accompagnement 60417 5008 Off power disconnection from the mains Éteint courant coupure avec le secteur 60417...

Страница 37: ... lamphouse cover Turn the lamphouse cover and remove it from the illumination unit Loosen the two clamping screw as illustrated below turn the clamping lever in direction of the arrow and remove the socket and bulb assembly Remove the old bulb use care as it may be extremely hot and insert the new bulb into the socket Place the bulb assembly into the housing by aligning the guide with the groove i...

Страница 38: ...sen the locking screw and remove the fixation target Do not over loosen the locking screw or it may drop Hold the top of the bulb and pull it out then insert the new bulb Insert the fixation target then tighten the locking screw Turn the power switch ON and check to see that the fixation bulb is properly illuminated Guide Locking screw Fixation target bulb ...

Страница 39: ...he same type of fuse as indicated in the holder T 2A 250V 100V 120V T 1A 250V 220V 240V REPLACING THE CHINREST PAPER If the chinrest paper supply is depleted remove the pins on the chinrest place the new pack age of paper over the chinrest and replace the two locating pins ADJUSTMENT OF THE SLIT WIDTH CONTROL KNOB If the movement of the slit width control knob becomes too light and the slit width ...

Страница 40: ...the standard accessories to remove the dust In case any dust still remains wipe it off using a soft cotton cloth moistened with a little alcohol Never use your finger or any hard object for cleaning 2 Cleaning the gliding plate base rail and shaft If the gliding plate or cross slide rail and shaft are dirty an unsmooth vertical or horizontal movement of the cross slide results Clean them by using ...

Страница 41: ...acement parts be sure to specify the product name part number and quantity required Product name Part number Appearance Illumination bulb 40340 20700 Socket 44613 25100 Chinrest paper 40310 40820 Fuse T 2A 250V 100V 120V T 1A 250V 220V 240V T24000030A T24000031A ...

Страница 42: ...ear the scale to enhance reading even in a darkened environment A magnifier is placed over the scale for easier reading 10 MEASURING EYEPIECE When this accessory is used in place of the normal eyepiece linear and angle measurements become possible It is used for the fitting of toric contact lenses as well Scale specifications Linear Scale 16mm 0 5mm minimum graduations Angle Scale 360 5 minimum gr...

Страница 43: ...ending on personal preference either the Haag Streit AG Model R 900 or Model T 900 could be used for measuring intraocular pressure AUTOMATIC INSTRUMENT TABLE AIT 15 Specifications Dimensions 510 W 450 D mm Table height 600 820mm Table size 490 500mm Weight approx 23kg Power consumption 270VA Source voltage 120V AC 50 60Hz USA and Canada 100 120V 220 240V AC 50 60Hz except for USA and Canada Model...

Страница 44: ...plate on the left of the base Period of Usage i e the purchase date Description of Problem as detailed as possible SLIT LAMP SL 3E INSTRUCTION MANUAL The 2005 version 2005 08 100TH Date of issue 1st August 2005 Published by TOPCON CORPORATION 75 1 Hasunuma cho Itabashi ku Tokyo 174 8580 Japan 2005 TOPCON CORPORATION ALL RIGHTS RESERVED 0 ...

Страница 45: ...100 Kuala Lumpur MALAYSIA Phone 03 42701192 Fax 03 42704508 TOPCON INSTRUMENTS MALAYSIA SDN BHD 77 162 Sinn Sathorn Tower 37th Fl Krungdhonburi Rd Klongtonsai Klongsarn Bangkok 10600 THAILAND Phone 440 1152 7 Fax 440 1158 TOPCON INSTRUMENTS THAILAND CO LTD Unit 18 4 Avenue of Americas Newington NSW 2127 AUSTRALIA Phone 02 8748 8777 Fax 02 9647 2926 www topcon com au TOPCON AUSTRALIA PTY LTD 2F Yoo...

Отзывы: