28
SPECIFICATIONS
SYSTEM CLASSIFICATION
Degree of protection against electric shocks: Type B applied part
Type B applied part is the applied part complying with the specified requirements of the
Standard IEC 60601-1 to provide protection against electric shock, particularly regarding
allowable LEAKAGE CURRENT.
Type of protection against electric shocks: Class
I
equipment
Class
I
equipment does not depend on basic insulation only for protection against electric
shocks. It can also be earthed; therefore, the metal parts with which one comes into contact
do not become conductive if the basic insulation fails.
The mode of operation: continuous operation equipment
Degree of protection against ingress of water: IP×0
SL-1E is the ordinary instrument (enclosed instrument without protection against ingress
of water)
Methods of sterilization or disinfection recommended by the manufacture: SL-1E do not have
any part to be sterilized or disinfected.
Not AP or APG equipment
PURPOSE OF USE
This instrument is used for the enlargement observation of eyeballs and the parts.
OPERATION PRINCIPLE
Illuminates the observed part by the illumination light emitted from the illumination optical sys-
tem and allows enlargement observation by binocular stereoscopic microscope.