59
Appendix
Specification
Measurement angle
:
2° only
Optics
Viewfinder field of view
:
5 °
Objective lens
:
f = 80mm F2.5
Photo detector
:
Photomultiplier Tube
Measurement distance
:
350mm
~
∞
(from the tip of the metal fixture on the
instrument of the objective lens)
Measurement Diameter
(Unit:
mmφ)
Measurement distance(
m
)
0.35
0.5
1
5
10
Diameter
10
15.4
32.8
169
341
※
The measurement distance is the distance from the tip of the metal fixture on the instrument
of the objective lens.
※
The diameter may change slightly according to the matching precision of the aperture mirror.
Measurement Function
The measured data are represented in following unit.
x
、
y
、
L
(x y : Chromaticity, L: Luminance)
u'
、
v'
、
L
(u' v' : Chromaticity, L : Luminance)
X Y Z
(X, Y, Z : Tristimulus values)
Tc
、
duv
、
L
(Tc : Color temperature, duv : Deviation)
Measurement time
About 1 second (SINGLE measurement, Y filter only, Range 4, Do not Auto calibration
)
Analogue output speed
80μs or less
※
The response speed in the table above is the time that it takes analog output from the
instrument to reach 90% of the peak value, when measuring an LED driven by a square
wave from a function generator.
Unit for Luminance
Candela per square meter (cd/m
2
)