55
4.3.30
LPG Command
This command instructs the instrument to send current LAN port number
Valid value range
:
49126 - 65535
PC
Command RD-80SA
“LPG”+CR+LF
“OK”+CR+LF
“50000”+CR+LF
“END”+CR+LF
4.3.31
PMT Command
This command sets sensitivity mode of photo detector
Valid value range
0
:
INITIAL
1
:
USER
PC
Command RD-80SA
“PMT”+CR+LF
“OK”+CR+LF
4.3.32
ERR Command
This command instructs the instrument to send latest error number
Error number
:
E0000 - E9999
PC
Command RD-80SA
“ERR”+CR+LF
“OK”+CR+LF
“E0003”+CR+LF
“END”+CR+LF