17 CHECK AND ADJUSTMENT
17.4 How to Set the Instrument Constant Value
To set the Instrument constant which is obtained in Section 17.1 “Check and adjusting of instrument
constant”, follow as below.
Operating procedure
Operation
Display
1
While pressing the [F1] key, turn power switch ON.
[F1]
+
POWER ON
2
Press the [F2] key.
[F2]
3
Enter the constant value. *1), 2)
Enter value
[F4]
4
Turn power switch OFF.
Power OFF
*1) Refer to Section 2.5 “How to Enter Alphanumeric characters”.
*2) To cancel the setting, press the [ESC] key.
ADJUSTMENT MODE 1/2
F1:V ANGLE 0 POINT
F2:INST. CONSTANT
F3:V0 AXIS P
↓
INST. CONSTANT SET
INST. CONSTANT
= - 0.6 mm
––– ––– [CLR][ENT]
INST. CONSTANT SET
INST. CONSTANT
: - 0.7 mm
INPUT ––– ––– ENTER