74
4.5 Error Code
When an error occurs in the instrument when connected with a PC and performing measurement,
the following error messages will be sent to the PC.
Error Code
Contents
E003
Displayed when there is an abnormal measurement angle. Make sure the
measurement angle is set properly.
E004
Displayed when a measurement command is sent before executing zero
adjustment. Execute zero adjustment.
E005
Displayed when time has elapsed since the previous calibration and it is
necessary to execute calibration by TOPCON TECHNOHOUSE factory.
E006
Displayed when the value of the correction factor is abnormal.
E007
Displayed when the value of the area correction factor is abnormal.
E008
Area correction limit write error.
Displayed when the value of one side of specified squared area is greater than
0.03.
E009
Area correction limit write error.
Displayed when specified area overlap with other area in same group.
E010
Area correction limit write error.
Displayed when the specified area does not place on the chromaticity diagram,
or when the maximum and minimum value of chromaticity area is improper.
E011
Area correction limit write error.
Displayed when the entered data does not correctly stored in instrument.
E012
Displayed when the single correction/direct correction dip switch setting and
communication command setting do not match. Make sure that the dip switch
setting matches to the communication command.
E013
Displayed when zero-adjustment is abnormal. There is a possibility that the
instrument do not make perfect dark condition during zero-adjustment, so
perform zero-adjustment with the eyepiece cap attached.
E
014
Displayed when the internal shutter is abnormal. There is a possibility that it
is broken, so contact the dealer or TOPCON TECHNOHOUSE.
E015
Displayed when average measurement is abnormal. There is a possibility that
the number of times that the instrument perform measurement are less than
necessary in average measurement mode, due to over range. Check the
measurement range setting for the measured object, and make sure the
measurement angle is properly set.
E016
Communication error.
The power OFF then back ON again.