Command Format: *STA 

Command Example:   

                                    Input Command: *STA 

                                    Instrument starts to output current 

 

*STO 

Function: Stop outputting current   

Command Format: *STO 

Command Example:   

                                    Input Command: *STO 

                                    If the instrument is outputting current, then stop outputting current 

4.2.2        Parameter Sub-System Command 

The parameter sub-system command is used to remotely set parameters.   

:PARA:CURR 

Function: Set or query the current current 

Command Format:   

                                Set Format: :PARA:CURR <data> 

                                Query Format: :PARA:CURR? 

Data:   

 

 

     

Data Type: Float, 4 bytes 

                                  Data Range

0~120 

                                  Data Accuracy: 0.1 

                                  Data Unit: A 

Set Example: To set the current to 1A, enter the command: :PARA:CURR 1 

            Query  Example:  If  you  enter  the  command:  :PARA:CURR?,  the  returned  content  is

1,  indicating  the 

current current is 1A. 

 

:PARA:FREQ 

Function: Set or query the current frequency 

Command Format:   

Set Format: :PARA:FREQ <data> 

                                Query Format: :PARA:FREQ? 

Data:   

 

 

     

Data Type: Float, 4 bytes 

                                  Data Range

0~2000000 

                                  Data Accuracy: 1 

                                  Data Unit: Hz 

            Set Example: To set the frequency to 100kHz, enter the command: :PARA:FREQ 100000 

            Query Example: If you enter the command: :PARA:FREQ?, the returned content is: 100000, indicating the 

current frequency is 100kHz.   

:PARA:FOOT 

Function: Set or query the current working state of the foot switch   

Command Format:   

Содержание TH1778A

Страница 1: ...1 TH1778A DC Bias Current Source Operation Mannual Tonghui Electronic Co Ltd www tonghui com cn...

Страница 2: ...Safety 6 1 4 9 Electromagnetic Compatibility 6 Chapter 2 Installation 7 2 1 Accessories 7 2 1 1 Current Output Clip 7 2 1 2 Footswitch 7 2 1 3 Link Line 7 2 1 4 Test Fixture 7 2 2 Front Panel Descript...

Страница 3: ...Output 18 Chapter 4 Command 19 4 1 TH1778A RS232 Instruction 19 4 2 SCPI Command 19 4 2 1 General command set 19 4 2 2 Parameter Sub System Command 20 4 2 3 State Sub System Command 21 4 2 4 System Su...

Страница 4: ...high precision and high frequency test needs The instrument has a new design friendly graphical operation interface and offers a variety of control modes manual key control foot switch control and re...

Страница 5: ...f forward current output If directional current is needed please change the connection direction between DUT and the output end Display the real time working state of the main machine and the slaver W...

Страница 6: ...power filter 1 4 6 Dimension W H D 430 185 473 mm3 1 4 7 Weight 18kg 1 4 8 Safety This tester is a Class I safe equipment 1 4 8 1 Insulation resistance In the reference working conditions the insulat...

Страница 7: ...ine is ET 07 2 and ET 54 Note when the current is greater than 40A please do not use it as the current output 2 1 2 Footswitch The standard footswitch is TH1801 001 2 1 3 Link Line Each TH1778S slave...

Страница 8: ...8 2 2 Front Panel Description 1 Power Switch Press down to power on push out to power off 2 Start Button To start output 3 Stop Button To stop output 4 DC Bias LED...

Страница 9: ...g the test signal and DUT produce the test access through AC coupling 13 Balance 14 Serial Port RS232 The output of the bias source and its parameters can be controlled through the serial port accordi...

Страница 10: ...utput current can be set on the setting interface or on the main interface Using the numeric keypad you can set the current value The default unit is A The seting range of current value is listed belo...

Страница 11: ...o the host and each slave and display in the table In addition to displaying the respective current it can also display the status of each slave and the host Status Meaning Off Indicates that the slav...

Страница 12: ...the numeric keyboard to set the response frequency the default unit is Hz The response frequency setting range is 0 2MHz 3 2 1 3 Rise Time Make use of the numeric keyboard to set the current rise tim...

Страница 13: ...se of the numeric keyboard to set the current step value the default unit is A the setting range is 0 120A 3 2 2 4 Scan Point Make use of the numeric keyboard to set the scan point amount the setting...

Страница 14: ...et the delay time 1 Move the cursor to the desired place and make use of the numeris keyboard to input each value 2 Move the cursor to the place where the delay time has been set already and make use...

Страница 15: ...on the switch to start outputting loosen the switch to stop outputting 3 3 1 3 Beep Volume To set the speaker to ON OFF 3 3 1 4 System Language The system language includes Chinese and English two op...

Страница 16: ...d there will be a progress bar to indicate the progress The instrument will restart automatically while the upgrade is completed 3 3 2 2 Restore Factory Setting Select Rest to restore factory setting...

Страница 17: ...r Serial Number of the instrument is provided 3 3 2 5 Password Users are able to set a new password or change the password 3 3 2 6 Slave Information You can use this feature to view the slave index an...

Страница 18: ...RT the instrument starts to output current 3 4 2 Stop Current Output 3 4 2 1 Stop Button Press the stop button to stop outputting current 3 4 2 2 Foot Switch In Trig mode step and release the foot swi...

Страница 19: ...ers can also buy the serial port cable from Tonghui 4 2 SCPI Command Note that all the following command are sent and received as strings Each command must be followed by a terminator or the instrumen...

Страница 20: ...ge 0 120 Data Accuracy 0 1 Data Unit A Set Example To set the current to 1A enter the command PARA CURR 1 Query Example If you enter the command PARA CURR the returned content is 1 indicating the curr...

Страница 21: ...ormat Query Format STAT HOST Returned Info Data type Unsigned integer 1 byte Data Range 0 31 0x1F Meaning Bit0 On or Off Bit1 Running or not Bit2 Over heat or not Bit3 Overload or not Bit4 Unbalanced...

Страница 22: ...g Query Example If you enter the command STAT WORK the returned content is running indicating the instrument is outputting current 4 2 4 System Sub System Command System Sub System Command is used to...

Страница 23: ...Chinese enter the command SYST LANG CHI Query Example If you enter the command SYST LANG the returned content is CHI indicating the current system language is Chinese SYST TRIG Function Set or query t...

Страница 24: ...nt output enter the command WORK START WORK STOP Function Stop current output Command Format WORK STOP Example To stop current output enter the command WORK STOP 4 2 6 Other Command REMO LOCK Function...

Отзывы: