Toho Electronics TSM-H1 Скачать руководство пользователя страница 1

 
 
                                                                                                                                 

   

 

Humidity-Voltage Converting Module TSM-H1

 

Operation Manual

 

1. Introduction

 

 

 

Thank you for purchasing this Relative Humidity-Voltage Converting Module TSM-H1. With the use of humidity sensor 

elements TD-A and TI-A, the product outputs relative humidity with 0 to 1V DC by simply connecting it to the power 
supply. Durability, high-speed response, and low hysteresis are some of the main characteristics of this product.   

 

 

2.    Wiring and Shape

 

    The wiring diagram of TSM-H1 is shown in Fig. 1. The shape of TSM-H1 is shown in Fig. 2. If the temperature sensor 
is selected as an option, its resistance value will be outputted with three wires.

 

 

3. Specifications

 

Specifications of TSM-H1 are the following:

 

Usable Temperature Range 

 

: - 25 to +80

 

Usable Humidity Range 

 

: 0 to 100

  RH

 

Relative Humidity Meter Measuring Accuracy : ±3

RH (5 to 95

RH at 25

 

Response Time (Humidity) 

 

: 15 sec. (If the membrane filter is attached: 90% response)

 

Power Voltage 

 

 

: 5 to 16V DC

 

Output (Humidity) 

 

 

: 0 to 1V DC (Output i

mpedance 100Ω)

 

(Temperature: Optional): Resistance (Three wires)

 

Dimensions 

 

 

:φ14mm × 

80mm (Excluding cable)

 

Cable Length 

 

 

: 1,000mm (Standard) 

 

Weight 

 

 

 

: Approx. 10g (Excluding cable)

 

Humidity Sensor Element 

 

:TD-A (Compatible humidity sensor element TI-A for TSM-H1-TIA)

 

Temperature Sensor Element (Optional): Pt100ΩCRZ2005, JIS class A

 

 

4. Chemical Resistance

 

    Using  TSM-H1  under  an  environment  that  contains  an  active  gas,  such  as  organic  gas,  may  cause  it  to  output  an 
abnormal value depending on the type and density of the active gas.

 

 

5. Replacement of Humidity Sensor (TSM-H1-TIA)

 

    In  case  an  abnormal  value  is  outputted  because  of  trouble  in  the  humidity  sensor,  replace  the  humidity  sensor 
element in accordance with the following procedures:

 

(1) Remove the sensor

’s

 protective cap by turning it. Remove the humidity sensor element.

 

(2)  Insert  the  new  sensor  and  then  attach  the  sensor

’s

  protective  cap.  Use  a  tweezer  to  avoid  touching  the  sensor 

surface.

 

 

No need to adjust the circuit after the replacement of the element.

 

 

6. Usage Precautions

 

    Please take note of the following upon using TSM-H1:

 

(1) Do not apply strong force on the product.

 

(2) Do not expose the product to direct sunlight.

 

(3) Do not expose the sensor and circuit sections to liquids such as water.

 

(4) Do not touch the humidity sensor section.

 

(5) Use the product under proper power voltage.

 

(6) Do not short-circuit the output.

 

(7) Do not wipe the product with an organic solvent.

 