Repeat steps a - b.
g. Disconnect the previously checked UV radiator and plug in another radiator that is
known to be working properly.
h. Put the lamp hoods on all muffs. Reboot the unit and check the UV radiator. Wait
2-3 minutes for UV radiator to warm up.
UV lamp is
NOT
working
Alarm system signals
failure
– red LED
Damaged electrical ballasts
Alarm system signals
proper work
– green
LED
Damaged electrical ballasts
and/or damaged alarm system
UV lamp is
working
Alarm system signals
failure
– red LED
Damaged alarm system
Alarm system signals
proper work
– green
LED
Everything works properly
13. TRANSPORT
The UV sterilizer should be shipped only in horizontal position. It should be disassembled for
shipping. The quartz sleeve and lamp should be packed separately as they are brittle. They
should be sufficiently well packed to withstand rough handling.