METINF005 
Thermometer 
Operating Instructions

The thermometer is a non-contact infrared 

HACCP thermometer that also includes a 

thermocouple probe. You can select only one 

measurement Mode at a time, but you can change 

the Mode at will. Please remember to keep away 

from baby and children and don’t use it for 

medically- related applications. 

* Special Features: 

  1. Food Code Zone Display; 

  2. Water Proof (IP65)

In Non-contact Infrared Thermometer function (IRT Mode)

# The white light wills automatically turns ON while the Scan button is pressed.

Simply aim the thermometer at the measure target with “Infrared Lens”  and press Scan (infrared) key to display the 

surface temperature. The distance to target ratio is 2.5:1 therefore the thermometer should be positioned as close to 

the target as possible.

Distance:Spot (FOV)= 2.5:1

Emissivity = 0.1~1 Step.01

Wave Length = 8-14um

While scanning, the newest temperature will be updated on the LCD  and the measurement will continue as long as 

the Scan (infrared) key is depressed. When the Scan (infrared) key is released, icon “Hold”  will appear on the display 

and the last measurement will remain visible for 15 seconds before the display goes blank.

Mode Selection   

 

MIN 

 MAX 

 LOCK 

 

o

C/ 

o

 EMIS

MINIMUM OR MAXIMUM MODE

The thermometer will display the minimum or maximum reading during the measurement period only until the Mode 

key is pressed.

To  utilize  the minimum  mode,  please  press Scan  (infrared)  key 

  Mode key 

 Scan  (infrared)  key.  And keep 

pressing Scan (infrared) key for measurement. 

To utilize the maximum mode, please press Scan (infrared) key 

 Mode key *twice

 Scan (infrared) key. And keep 

pressing Scan (infrared) key for measurement.

LOCK MODE

The lock mode is particularly useful for continuous monitoring of  temperatures. The thermometer  will continuously 

display the temperature for up to 60 minutes or until the Scan (infrared) key button is pressed.

To utilize the lock mode, please press Scan (infrared) key 

 Mode key *three times

 Scan (infrared) key. 

o

C OR 

o

F MODE

To change the ‘

o

C’ or ‘

o

F’ mode, please press Scan (infrared) key 

 Mode key *four times

 Scan (infrared) key.

Same steps can be taken when switching from 

o

F to 

o

C.

EMISSIVITY

The infrared thermometer is supplied with a default emissivity of 0.95. The emissivity can be changed from 0.10 

(10E) to 1 (100E). Changes should only be carried out by experienced personnel. For information relating to the 

emissivity of specific materials, please contact the nearest retailer. Note: non-contact infrared thermometers are 

not recommended for use in measuring the temperature of shiny or polished metals.

To change the emissivity, please Scan (infrared) key 

 Mode key *five times

 Scan (infrared) key for 

each 0.01 (1E) adjustment

 Mode key.

In Contact Thermocouple Probe function (COT Mode)

Attach the thermometer at the measure target with “Probe” and press Probe key to continuously display the 

temperature for up to 4 minutes. After that the device will automatically shut off to extend the battery life. 

Press Probe key will interrupt the scanning to display the last temperature with a 'Hold' wording. 

To reenter scanning just press Probe key again.

 1. Do not twist the probe and rotate the probe in wrong direction.

2. Over stress on probe may cause break.

3. After measure high temp, the probe may remain HOT for a while.

4. Probe is dangerous for human when the probe is in an open position. Remember to hold the probe

back when not in use. 

 The probe of contact thermometer may be damaged if

 

exceeding the specification of

    measurement temperature range.

 

To avoid electric shock and thermometer damage, do not measure live circuit where voltage 

   exceeding 24V AC RMS or 60V DC

 

with the thermocouple probe.

HACCP check

The  “HACCP  CHECK”   feature  is  incorporated  in  our  thermometer 

temperature to  graphically indicate  critical  temperature  zone.  The icons 

and  LED  indicators  located  above  the  display indicate  a  food  product 

stays in a safe or unsafe HACCP “  Danger Zone”  temperature. The green 

and red LED light will always be lit before power off. 

A Green  LED  appears with icon  ”     ”  indicates a  safe  cool or  frozen 
condition  below  4

o

C(40

o

F)  or  appears with icon  ”     ”  indicates  a  safe 

holding temperature above 60

o

C(140

o

F).

When temperature is between 4

o

C and 60

o

C, the red LED  with icon ”    

”  will appear and indicate that the temperature is fallen within the HACCP 
“Danger Zone” from 4

o

C to 60

o

C (40~140

o

F)

.

LCD ERROR MESSAGES

The thermometer incorporates visual diagnostic messages as follows:

 

 

‘Hi’ or ’Lo’ is displayed when the temperature being measured is outside of the 
range of the instrument, ‘Hi’ when higher than +250

o

C (482

o

F) and ‘Lo’ when 

lower than -55

o

C (-67

 o

F).

‘Er2’ is displayed when the thermometer is exposed to rapid changes in the 
ambient temperature. ‘Er3’ is displayed when the ambient temperature exceeds 0

o

(32

o

F) or +50

 o

C (122

 o

F). The thermometer should be allowed plenty of time 

(minimum 30 minutes) to stabilize to the working/room temperature.

For all other  error messages it is necessary to reset the thermometer. To reset it, waiting for auto 
power off

,

 remove the battery and wait for a minimum of one minute, reinsert the battery and turn 

on. If the error message remains please contact the Service Department for further assistance.

For all other  error messages it is necessary to reset the thermometer. To reset it, waiting for auto 
power off

,

 remove the battery and wait for a minimum of one minute, reinsert the battery and turn 

on. If the error message remains please contact the Service Department for further assistance.

BATTERIES

STORAGE & CLEANING

The sensor lens is the most delicate part of the thermometer. The lens should be kept clean at all times, care should 

be taken when cleaning the lens using only a soft cloth or cotton swab with water or medical alcohol, allowing the lens 

to fully dry before using it, do not submerge any part of the thermometer. The  thermometer  should be stored at room 
temperature between –20 to +65

 o

C (-4 to 149

 o

F).