Technical data
XD 7500
162
ba77191e01 - 71310060
04/2018
7
Technical data
7.1
Measurement characteristics
Measuring principle
Spectrophotometer with reference beam technology
Light source
Lamp type
Xenon flashlamp
Average lifetime
5 x 10
8
flashes, corresponding to at least
13000 h in permanent operation
Monochromator
Type
Grating monochromator with step motor
Wavelength range
190 - 1100 nm
Max. scan speed
approx. 1000 nm/min
Wavelengths calibration
Automatic
Accuracy
± 1 nm
Reproducibility
± 0,5 nm
(can be checked, e. g. with Holmium oxide fil-
ter)
Resolution
1 nm
Spectral band width
4 nm
Photometric mea-
surement
Light sensor
Photo diode
Measuring range
A = -3.300 to A = +3.300
Linearity
< 1 % for E
≤
2,000
in the range from 340 to 900 nm
Accuracy
± 0.003 A for A < 0.600
± 0.5 % of the reading for 0.600
≤
A
≤
2.000
Reproducibility ±
0.002 at A = 1.000
Resolution
Δ
A = 0.001
Scattered light
< % Transmission at 340 and 408 nm
*
in the range from 200 nm to 1000 nm