4 Preparation
4-12
MIRUS Controller, SW 2.03.02, Additional Information Master (en-GB)
If the MIRUS Reflector is not parked in the
Park Bay the following message appears on
the screen.
During System test the MIRUS Controller
starts the operation of the heating system and
checks the following:
•
MIRUS Reflector pneumatic integrity for
leaks and kinked hoses.
•
MIRUS Controller valve pneumatic
integrity.
4.7 Anomaly in start-up sequence
In rare cases it can lead to an anomaly in
start-up sequence.
To restart MIRUS Controller
•
disconnect mains supply.
•
wait for 5 seconds (green lights off).
•
reconnect mains supply.
•
restart by pressing the Confirm button
(usual start).
HW 2.n
SW 2.03.02
System test failed!
Reflector not in Park Bay.
Please insert the Reflector
and repeat the System test.
Sorry. A problem has been detected.
The device cannot be operated.
Please remove AC supply and
wait until all device lights are off.
Then reconnect AC supply and
restart device.
Содержание Mirus MC DES
Страница 1: ...Additional Information MIRUSTM Controller Software version 2 03 02 Hardware version 2 n...
Страница 28: ...1 Introduction 1 22 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 48: ...3 General description 3 8 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 78: ...5 Operation 5 18 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 92: ...6 Cleaning and maintenance 6 14 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 118: ...7 Alarms and messages 7 26 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...