background image

  Air Velocity 

Range 

Resolution 

Accuracy

 

m/s 

0.40—30.00 

0.01 

±3%  ±0.20 m/s

 

ft/min 

80—5900 

±3%  ±40 ft/min

 

km/h 

1.4—108.0 

0.1 

±3%  ±0.8 km/h

 

MPH 

0.9—67.0 

0.1 

±3%  ±0.4 MPH

 

Knots 

0.8—58.0 

0.1 

±3%  ±0.4 knots

 

Air Flow 

Range 

Resolution 

Area

 

CFM 

0-999900 

0.001 

0-999.9 ft

2

 

CMM 

0-999900 

0.001 

0-999.9 m

2

 Air Temperature 

Range 

Resolution 

Accuracy

 

 

14-140°F (-10-60°C) 

0.1°F/C 

±4.0°F (2.0°C)

 

Infrared 

Range 

Resolution 

Accuracy

  Temperature

 

 

-58 to -4°F (-50 to -20 °C) 

0.1°F/C 

±9.0°F (5.0°C)

 

 

-4 to 932°F (-20 to 500°C) 

0.1°F/C 

±2% reading or ±4°F (2°C)

SPECIFICATIONS

Back sheet 1 of 

2

 

© 2008 SPX

CFM(ft

3

/min)=Air Velocity(ft/min)XArea(ft

2

)

CMM(m

3

/min)=Air Velocity(m/s)XArea(m

2

)X60

CFM: cubic feet per minute
CMM: cubic meters per minute

 

 

m/s 

ft/min 

km/h 

MPH 

knot

 

1 m/s 

196.87 

3.60 

2.24 

1.944

  1 ft/min 

0.00508 

0.01829 

0.01138 

0.00987

 

1 km/h 

0.2778 

54.69 

0.6222 

0.54

 

1 MPH 

0.4464 

87.89 

1.6071 

0.8679

 

1 knot 

0.5144 

101.27 

1.8519 

1.1523 

1

     AVG

  =  Press to obtain an average reading for 

 

  multiple points in Flow mode or Velocity

 

  mode. Up to 20 points can be aver

 

  aged.

   AREA

  =  Press and hold to enter into the Area 

 

  mode (ft

2

 or m

2

) in CFM (ft

3

/min) or 

 

  CMM (m

3

/min) modes. When recording 

 

  MAX/MIN/AVG readings for a single 

 

  point, press the Area button to clear 

 

  previous readings.

   HOLD

  =  Press once to freeze the displayed 

 

  reading. Press again to unlock display.

 

=  Press to turn ON the backlight. Press 

 

  again to turn OFF.

MAX/MIN (TEMPERATURE)

  =  Press to record 

and store the highest and lowest readings for probe 

temperature.

HOLD (TEMPERATURE)

  =  Press to freeze the 

displayed reading for probe temperature. Press again 
to unlock display.

KEYPAD BUTTONS CONT’D