3.3.2 Operation with other Supply and Evaluation Units
The transducer can be operated with other customer related display units or volta-
ge supplies.
The electrical connection is to be made by means of suitable cables considering
EMI demands and according to the pin description shown below:
Socket
Sub-D, 15-pole, male
Pin1:
Identification: 15 k
Ω
Pin2:
Signal Output 0-10 VDC
Pin3:
AGND
Pin4:
Voltage Supply 24 VDC
Pin5:
Supply GND
Pin6:
Relay 1, N.O.
Pin7:
Relay 1, Common
Pin8:
Relay 1, N.C.
Pin9:
-
Pin10:
RS485 +
Pin11:
RS485 -
Pin12:
Shield
Pin13:
Relay 2, N.O.
Pin14:
Relay 2, Common
Pin15:
Relay 2, N.C.
We recommend to have Shield (pin 12) and supply common (pin
5) grounded in the supply unit.
Incorrect connection or inadmissible supply voltage can damage
the transducer!