Important:
All pressure MUST be vented before disconnecting the air supply
hose or personal injury may result.
7
INSPECTION and MAINTENANCE
7.1
Recommended Maintenance Schedule
Periodic
Hours
Maintenance
100% Duty Cycle
Initial Inlet/Outlet Filter Inspection
1
st
500
To establish service period
Minor Service Kit
1,500
Major Pump Rebuild or
Replacement
3,000
7.2
Filter Inspection
The Intake Filter, Outlet Filter, and Regulator Muffler require periodic inspection. Initial
inspection is suggested at 500 hours; then user should determine the frequency
thereafter. Most problems can be prevented by keeping the intake filter and particulate
filter clean. A dirty intake filter will decrease pump performance and can decrease pump
life.
Warning
: Disconnect power and be sure all pressure has
been vented before service!
1. Intake
a. Remove the snap-fit cover.
Filter (LF1) b. Clean felt filter using air or vacuum.
c. Replace snap-fit cover.
2. Particulate a. Remove the hose by turning the flare nut counterclockwise.
Filter (LF2) b. Unscrew the adapter (1) by turning counterclockwise.
c. Remove filter (3) and clean with water and blow dry or replace.
d. Inspect bowl (4 & 5) and seal (2). If dirty, clean by wiping
the bowl with a soft dry cloth.
e. Insure seal is clean and in place then re-install filter and
rotate the adapter clockwise until snug.
f. Re-install hose flare nut using a backup wrench turning
clockwise.
Particulate Filter (LF2)
Thunder Scientific Corporation ACS-517B Air Compressor System
1-8
Document Number: ACS-517B.doc - Edition 06 - April 2014