Thumper

 Model 1007

CCV

  

Mechanical CPR System

 

Operation Manual 

 

 

 

 

 

(Part Number 14799-02) 

 
 
 
 
 
 
 
 
 

 

 
 
 
 

 

     

 

 

Manufactured in the USA by:

 

 

 

 PROTECTED UNDER ONE    

         

                

4717 Talon Court SE 

       

OR MORE OF THE 

                     Grand Rapids, MI  49512   USA

 

FOLLOWING U.S. PATENTS:

          Tel:  (800) 530 9939 or (616) 554 9696     

 

 

    6,171,267     5,743,864

    

 

       Fax:  (616) 554 3067 

        e-mail:  [email protected] 

        website:  

www.michiganinstruments.com

 

 

 

Содержание 1007CCV

Страница 1: ...er 14799 02 Manufactured in the USA by PROTECTED UNDER ONE 4717 Talon Court SE OR MORE OF THE Grand Rapids MI 49512 USA FOLLOWING U S PATENTS Tel 800 530 9939 or 616 554 9696 6 171 267 5 743 864 Fax 6...

Страница 2: ...REV 07 11 2 Blank Page Intentional...

Страница 3: ...ling 13 C SETUP AND OPERATION Precautions 17 Recommendations 17 Positioning the Patient 18 Thumper Model 1007CCV Deployment 19 To Remove the Thumper from the Patient 24 D STORAGE AND SHIPPING Storage...

Страница 4: ...REV 07 11 4 Blank Page Intentional...

Страница 5: ...provide CPR support under the direction and control of a licensed physician Use of this device for any other purpose is strongly discouraged CONTRAINDICATION There are situations where CPR is not the...

Страница 6: ...is manual appropriate training and adequate practice with the device This manual contains important information on all aspects of operating and maintaining the device After a complete review use it as...

Страница 7: ...follow the instructions for preventive maintenance and cleaning procedures after each use They are found in the Care Cleaning and Maintenance section of this manual Caution Submersion of the Thumper...

Страница 8: ...REV 07 11 8 Blank Page Intentional...

Страница 9: ...CV is a mechanical automatic CPR device that can be set up in seconds It is powered by compressed oxygen and is electrically insulated allowing it to be freely and safely used in conjunction with rout...

Страница 10: ...tes an oxygen powered time cycled constant flow ventilator commonly known as a Patient Demand Valve PDV to deliver ventilation Tidal volume can be set from 400ml to 1200ml with delivery synchronized t...

Страница 11: ...to retain the oxygen during a source change Wall adapters are available that connect to the O2 Supply Hose to allow connection to the various and most common hospital and ambulance oxygen pipeline sys...

Страница 12: ...able separately Figure 3 Mobile Oxygen Carriers MOC The Carrying Storage Case The carrying storage case is constructed of a durable Cordura 1000 denier nylon The Thumper BackBoard O2 Supply Hose and c...

Страница 13: ...humper Model 1007CCV s controls are conveniently located in one area to assist the user Refer to the following illustration for Figure references to the controls and labeling described Figure 5 Figure...

Страница 14: ...1 This control allows the operator to turn on RUN chest compressions or turn off STOP chest compressions The system is controlled by pressing in and rotating the switch to the desired position RUN or...

Страница 15: ...gle switch selects the ventilation mode of the Thumper Model 1007CCV ON or OFF The arrows on the label indicate the switch position for the desired mode With the switch in the up or ON position ventil...

Страница 16: ...REV 07 11 16 Blank Page Intentional...

Страница 17: ...the Thumper in operation ensure the BackBoard ensure the patient is secured snugly to the BackBoard using the provided shoulder straps Also ensure the patient is properly secured to the stretcher as...

Страница 18: ...nt in cases of suspected C spine injury always support the patient s head in a neutral position 2 Place the BackBoard under the patient orienting the head of the patient in the direction indicated by...

Страница 19: ...sed to the PDV E Ensure the oxygen source is energized then attach the O2 Supply Hose to the oxygen source first Pull the collar back from the O2 Source end connector press firmly onto the male connec...

Страница 20: ...mper with one hand at the Base of the Column and the other by the Handle B Insert the Base into the side slot of the Backboard on whichever side of the patient is most convenient Slide the Base into t...

Страница 21: ...y referring to the Sternal Deflection Number located on the scale on the back of the Column The blue arrows indicate the depth required to provide the 20 A P sternal deflection for the patient Warning...

Страница 22: ...trol 1 to RUN D Rotate COMPRESSION DEPTH Control 2 slowly clockwise until sufficient compression depth is demonstrated by viewing the Piston at eye level Increase the control until the top of the Pist...

Страница 23: ...l 3 is set too high or inadvertently bumped Warning Patient ventilation may be insufficient if Ventilation Volume Control 3 is set too low or inadvertently bumped Warning Should the blue rubber diaphr...

Страница 24: ...al interruption of the cardiac compressions as described herein may be required to avoid the ECG motion artifact associated with cardiac compressions The Thumper Model 1007CCV is electrically insulate...

Страница 25: ...the BackBoard I Remove the BackBoard from the patient J Clean and inspect the Thumper per the recommended Shift Check Daily or after each use Warning The Breathing Hose PN 14669 Non Re breathing Valv...

Страница 26: ...REV 07 11 26 Blank Page Intentional...

Страница 27: ...uick access to the Thumper at an emergency site A Thumper that is stored assembled should be placed on a crash cart or other surface where it will be used The Arm should be positioned at the top of th...

Страница 28: ...REV 07 11 28 Blank Page Intentional...

Страница 29: ...ting the Patient Demand Valve section below Disinfection Guidelines Standard colorless chemical disinfectant solutions may be used to wipe down external surfaces of the Thumper device O2 Supply Hose B...

Страница 30: ...ten snugly Check the Lip Valve to assure the flapper is not twisted Properly align the locating bosses and set the Lip Valve inside the Pressure Limit Alarm Install the Hose Adaptor on the Pressure Li...

Страница 31: ...all parts thoroughly before reassembling Inspect all parts for damage or breakage and replace if needed Place the lip valve into the adaptor ensuring the lip points to the outlet of the adaptor patie...

Страница 32: ...utine maintenance of internal components performed at the manufacturer by factory trained personnel The schedule for performing these procedures should be determined by the user taking into considerat...

Страница 33: ...ks cracks cuts worn hose or damaged connectors Connect the O2 Supply Hose to the oxygen source first then connect it to the Thumper 6 Verify that the device cycles when the O2 Supply Hose is connected...

Страница 34: ...signs of wear 6 Device cycles and Pressure Indicator shows green 7 Ventilation delivered at 600ml Pressure Limit Alarm activates 8 No ventilation with Vent Mode OFF Control 3 rotates and secure 9 O2...

Страница 35: ...n the Arm Lock Lever Attach the O2 Supply Hose to the oxygen source first b Check the operation of the O2 Supply Hose by connecting and disconnecting the Thumper end of the supply hose a few times The...

Страница 36: ...t smoothly in both BackBoard slots 4 Ventilation Compression test a Piston reads and Arm Lock Lever secures Arm in place b O2 Supply Hose connects and disconnects easily c Device cycles and Pressure I...

Страница 37: ...t secured to valve shaft Verify Control 3 knob is secured to shaft Caution If Control 3 is found loose on the shaft and subsequently tightened the indicator on the knob may not be properly aligned to...

Страница 38: ...n 60 0 msec Ventilation Time cycled constant flow ventilator Calibrated Volume Control 400 1200 100 mL or 10 adjustable from 400 1200 mL 0 105 0 317 0 026 gal or 10 adjustable 0 0 317 gal Deliver 95 5...

Страница 39: ...acement Cap Hose Adaptor Oxygen Management 15290 MOC Dual D Tank Soft Case 15300 MOC Dual E Tank Soft Case 11732 Aluminum Oxygen Tank D size 11733 Aluminum Oxygen Tank E size 15040 Soft Case for Carbo...

Страница 40: ...REV 07 11 40 Blank Page Intentional...

Страница 41: ...ches or dents which may result from normal use of this equipment or malfunctions due to mishandling or improper packaging The Thumper is designed to be used during resuscitation and transportation of...

Страница 42: ...ts or any service related questions should also be directed to the Service Department C If your Thumper CPR System must be returned to Michigan Instruments Inc please observe the following procedures...

Страница 43: ...evaluation fee plus return shipping charges This fee will be charged ONLY if repairs are not authorized and the device must be returned unrepaired We are obligated to label and tag as unusable any Th...

Страница 44: ...REV 07 11 44 Blank Page Intentional...

Страница 45: ...d by 1 Lack of breathing 2 Lack of pulse and heart sounds ECG Electrocardiogram A graphic tracing of the electrical current caused by contraction of the heart muscle EMS Emergency Medical Service MOC...