The three threshold switches,
LOCK/OFF
,
CUE TONE/OFF
and
BLW/ABV
are only functional when the front panel power switch
is set to
THR
. Possible combinations are:
LOCK/OFF
CUE/OFF
ABV/BLW
1
OFF
OFF
ABV
Proportional tone feedback above threshold.
2
OFF
OFF
BLW
Proportional tone feedback below threshold.
3
OFF
CUE TONE
4 or 60 sec.
ABV
Cue tone when EMG above threshold for >
4 or 60 seconds. (Goes off immediately
when EMG goes below threshold).
4
OFF
CUE TONE
BLW
Not a valid position. Cue tone (CUE TONE)
works only for above (ABV) threshold
position.
5
LOCK
CUE TONE
ABV
Same as condition 3; however the tone
locks on until the power switch
(OFF/CONT/THR) is turned off.
6
LOCK
CUE TONE
BLW
Not a valid position.
7
LOCK
OFF
ABV
Not a valid position.
8
LOCK
OFF
BLW
Not a valid position.
Side Panel Jacks
(see figure 8)
INPUT:
Plug the MyoScan sensor in here.
Plugging the earphones in here disconnects the internal
speaker and sends tone feedback through the earphones.
0-2V:
Provides a 0-2 Volts output signal. The MyoTrac can be
connected to other third party device with an isolated voltage input
that accepts 0-2V.
Make certain the instrument complies with IEC
601-1.
Note: This output is the same for all the gain settings. An output
value of 2 Volts will represent 20
V at the
x1
setting, 200
V at
x10
and 2000
V at
x100
.