© 2007 - 2018 Thorlabs
44
Wavefront Sensor
Note
The accuracy of the calibration increases if image averaging is set to > 1, see
panel.
The generated user calibration data are only valid for the actual MLA and camera image size,
that are selected in the
and
panel. Switching to
an other MLA or camera resolution will require a new calibration!
If the '
User Calibration
' option is enabled, the stored file for a particular camera, MLA, and res-
olution is automatically loaded at program start and every time the camera resolution or MLA is
changed.
5.2.1.4.1 Spherical Wavefront
User Calibration using a Spherical Wavefront
In order to perform a spherical wavefront user calibration you need to apply a high-quality diver-
gent beam that originates from a point source with a perfect spherical wavefront. A single-mode
fiber output without any collimators or lenses is ideally suited for that purpose. Be sure the
beam is properly aligned to the Wavefront Sensor and fills the selected active camera area
completely.
Note
The calibration fails for insufficiently illuminated regions of the camera sensor. Although no er-
ror occurs, the corresponding area will be excluded from further wavefront measurements.
1. Select the
Internal Reference
in the
:
2. Adjust the point source so that it is centered in front of the Wavefront Sensor at a distance in
the range 100 mm to a maximum of 5 m.
Note
The geater the distance between the point source and the WFS, the lower the
curvature of the wavefront. In other words, the spot displacements decrease and the meas-
urement of the wavefront curvature becomes more sensitive to noise and ambient light. In
addition, the power density on the CMOS sensors drops with the square of distance, which
also increases the impact of noise sensitivity on the measurement. For these reasons, it is
recommended that the distance to the source not exceed 500 mm.
The Radius of Curvature (RoC) is calculated and displayed by the WFS software. Assuming
a point source and propagation in air, the RoC is equivalent to the distance between the
camera sensor and the light source.
3. Start the calibration procedure using
Menu
à
Calibration
à
Spherical Wavefront
or
simply click on the appropriate symbol
in the toolbar. The dialog window illustrates the re-
quired calibration setup:
25
25
35
24
Содержание WFS20-14AR
Страница 2: ...Copyright 2007 2018 Thorlabs Version Date 5 0 19 Jul 2018...
Страница 15: ...2007 2018 Thorlabs 3 Coordinate Definitions 13 Wavefront Graph...
Страница 99: ...2007 2018 Thorlabs 8 Appendix 97 Typical Wavelength Response Typical Response of the WFS20 CMOS Camera...
Страница 106: ...2007 2018 Thorlabs 104 Wavefront Sensor Selectable camera image sizes for WFS40 Normal Mode...
Страница 107: ...2007 2018 Thorlabs 8 Appendix 105 Selectable camera image sizes for WFS40 sub2 Mode...
Страница 123: ...2007 2018 Thorlabs 8 Appendix 121 8 9 2 Drawing WFS20 5C...
Страница 124: ...2007 2018 Thorlabs 122 Wavefront Sensor 8 9 3 Drawing WFS20 5C M...
Страница 125: ...2007 2018 Thorlabs 8 Appendix 123 8 9 4 Drawing WFS20 7AR...
Страница 126: ...2007 2018 Thorlabs 124 Wavefront Sensor 8 9 5 Drawing WFS20 7AR M...
Страница 127: ...2007 2018 Thorlabs 8 Appendix 125 8 9 6 Drawing WFS20 14AR...
Страница 128: ...2007 2018 Thorlabs 126 Wavefront Sensor 8 9 7 Drawing WFS20 14AR M...
Страница 129: ...2007 2018 Thorlabs 8 Appendix 127 8 9 8 Drawing WFS20 Control Box...
Страница 130: ...2007 2018 Thorlabs 128 Wavefront Sensor 8 9 9 Drawing WFS30 5C...
Страница 131: ...2007 2018 Thorlabs 8 Appendix 129 8 9 10 Drawing WFS30 5C M...
Страница 132: ...2007 2018 Thorlabs 130 Wavefront Sensor 8 9 11 Drawing WFS30 7AR...
Страница 133: ...2007 2018 Thorlabs 8 Appendix 131 8 9 12 Drawing WFS30 7AR M...
Страница 134: ...2007 2018 Thorlabs 132 Wavefront Sensor 8 9 13 Drawing WFS30 14AR...
Страница 135: ...2007 2018 Thorlabs 8 Appendix 133 8 9 14 Drawing WFS30 14AR M...
Страница 136: ...2007 2018 Thorlabs 134 Wavefront Sensor 8 9 15 Drawing WFS40 5C...
Страница 137: ...2007 2018 Thorlabs 8 Appendix 135 8 9 16 Drawing WFS40 5C M...
Страница 138: ...2007 2018 Thorlabs 136 Wavefront Sensor 8 9 17 Drawing WFS40 7AR...
Страница 139: ...2007 2018 Thorlabs 8 Appendix 137 8 9 18 Drawing WFS40 7AR M...
Страница 140: ...2007 2018 Thorlabs 138 Wavefront Sensor 8 9 19 Drawing WFS40 14AR...
Страница 141: ...2007 2018 Thorlabs 8 Appendix 139 8 9 20 Drawing WFS40 14AR M...
Страница 144: ...2007 2018 Thorlabs 142 Wavefront Sensor 8 11 Certifications and Compliances...
Страница 145: ...2007 2018 Thorlabs 8 Appendix 143...
Страница 153: ...www thorlabs com...