Copyright © 2017 Thorlabs
6
TED4015
This product has been tested and found to comply with the limits according to
IEC 61326-1 for using connection cables shorter than 3 meters (9.8 feet).
Vous pouvez trouver les traductions françaises des paragraphes ayant trait à la sécurité
d'utilisation de ce produit sur le lien suivant:
https://www.thorlabs.com/_sd.cfm?fileName=18861-D04.pdf&partNumber=TED4015
En outre, vous pouvez soit scanner le QR code, soit vous référer à la section "Docu-
ments" sur la page web du produit.
1.2 Protection of the TEC Element
To protect the connected TEC element (Peltier element) the TED4015 Temperature Controller
includes the following protective circuits:
Limit of the TEC current in all operating modes
Protection against thermal destruction.
Sensor fault detection
Protection against use of incorrect temperature sensors / protection against line interruption of
the temperature sensor.
Contact protection of the TEC element (open circuit)
Protection against cable damage, bad contact or too high TEC element resistance.
Control LED for TEC current ON
Protection against accidental turning off the TEC output.
Over-temperature protection
Protection against malfunction caused by internal overheating of the controller.
Defined states after switch-on
When switching on the TED4015 Temperature Controller with the mains switch the output is al-
ways switched off.
Mains filter
Protection against line bursts or transients.
Line failure monitoring
In case of line failure / line interruption the instrument will wake up anew as if it has been
switched on. So the output remains switched off.