Copyright © 2017 Thorlabs
46
TED4015
TED4015
Front Panel
Remote Control
Over temperature protection
yes
Short circuit when TEC off
yes
Temperature setpoint limit
yes
Temperature window protection
yes
TEC current limit
yes
TEC open circuit protection
yes
General Data
Line voltage
100 to 120V ±10% / 200 to 240V ±10%
Line frequency
50 to 60 Hz ±5%
Maximum power consumption
600 VA
Supply mains over-voltage
Category II (Cat II)
Operating temperature (non condensing)
0 to +40°C
Storage temperature
-40 to 70°C
Relative humidity
max. 80% up to 31°C, decreasing to 50% at 40°C
Pollution degree (indoor use only)
2
Operation altitude
< 2000 m
Warm-up time for rated accuracy
30 min
Weight
5.3 kg
Dimensions (WxHxD) without operating elements
263 x 122 x 307 mm³
Dimensions (WxHxD) with operating elements
263 x 122 x 345 mm³
1
control range and thermal stability dependent on thermistor parameters
2
depending on the selected measurement range
All technical data are valid at 23 ± 5°C and 45 ± 15% rel. humidity (non condensing)