3 Operating Instructions
© 2019 Thorlabs
19
the actual SPCM status in the status bar, in this child window single error events are recor-
ded until a new measurement is started.
·
Status bar
displays the measurement and connect status.
·
Graphical and auxiliary displays
are arranged in 4 tabs
3.4.6.1 Alignment Display
This display is useful to adjust the optical setup. Beside the actual photon count it brings up im-
portant information on the current measurement and device status.
Measurement State
halted / running. Controlled by Start / Stop button
Counter Gate State
active (=counting) / not active
Trigger State
disabled (Manual and Free Running Timed Counter modes)
enabled (all triggered modes)
Temperature of APD
"over temperature": temperature exceeds set point for > 0.5 K (see
)
Saturation of APD
(see
)
Counter overflow
"overflow" indicates that more than 2
31
-1 = 2,147,483,647 photons
were counted
USB Overload
"overload" indicates that measurement speed exceeds the USB
transfer rate.
51
49
Содержание SPCM A Series
Страница 1: ...Single Photon Counting Module SPCMxxA Operation Manual 2019 ...
Страница 2: ...Version Date 1 2 27 May 2019 ...
Страница 11: ...2 Installation 2019 Thorlabs 9 After completing the installation you might be prompted to restart your PC ...
Страница 42: ...40 SPCMxxA 2019 Thorlabs ...
Страница 43: ...5 Maintenance and Service 2019 Thorlabs 41 ...
Страница 52: ...50 SPCMxxA 2019 Thorlabs 6 3 Pulse Out Below a typical pulse shape at the Pulse Out SMA connector is shown ...
Страница 54: ...52 SPCMxxA 2019 Thorlabs 6 5 Dimensions Dimensions of the SPCMxxA imperial version ...
Страница 55: ...6 Appendix 2019 Thorlabs 53 Dimensions of the SPCMxxA M metric version ...
Страница 61: ...www thorlabs com ...