background image

© 2019 Thorlabs

13

PM102x

3.2.7   UART Operation

The PM102 is delivered in RS232 operation but can be switched to UART operation. 

UART dependent PM102 functions:

The PM102 may need to be switched from RS232 to UART operation when using an external
micro controller for PM102 control. In RS232 operation the PM102 uses voltage levels of +/-5
V, while in UART mode the voltage level is shifted to 0/5 V. The PM102 can also be adjusted to
accommodate micro-controllers using 3.3 V. When switched to 3.3 V, UART operation is toler-
ant for 5 V input signal.

For instructions on how to change the voltage level for UART operation, please refer to the doc-
ument Write-Your-Own-Applications (WYOA) for PM102x. 

For  UART  operation,  please  use  standard  protocols  for  UART  communication  and  the  SCPI
commands listed in WYOA for PM102x.

3.2.8   Reset Button

This button reboots the internal micro controller of the PM102x. To press the button, please use
a  small  device  that  will  not  break  inside  the  instrument.    All  settings  are  saved  due  to  the  in-
ternal non-volatile memory. 

3.2.9   Baud Rate

For PM102, the Baud rate can be adjusted  from 9600 to 230400 Bit/s. This can be done either
via  the  USB  interface  or  via  the  serial  interface  using  SCPI  commands  or  driver  functions.
Please see the document Write-Your-Own-Applications (WYOA) for PM102x. The default baud
Rate is 115200. 

 Note 

After adjusting the Baud Rate via RS232 or UART, close the RS232 or UART connection and
reconnect  with  the  desired  Baud  Rate  in  order  for  the  Baud  Rate  to  communicate  it  via
RS232/UART. This is not required for operation via USB. 

3.2.10   Non-Volatile Memory

All  PM102x  models  have  an  internal  non-volatile  memory  for  all  settings  (i.e.  wavelength  set-
ting, ranging, analog output configuration). The PM102x will start in the same configuration as
prior to shut-down. 

3.2.11   Auxiliary I/O Ports

PM102 with its DA-15 connector gives access to two configurable general purpose input/output
(I/O) Pins (DA-15 Pins 2 and 3). Please see the 

Pin assignment

 in the appendix. These aux-

iliary I/O ports use 3 V logic for output and are tolerant for 5 V input signal.

The  I/O  port  can  be  controlled  or  monitored  via  the  instrument  driver  or  through  SCPI  com-
mands in custom made software.

19

Содержание PM102

Страница 1: ...Optical Power Meter Interfaces 2019 PM102 PM102A PM102U Operation Manual...

Страница 2: ...Version Date 1 0 11 Dec 2019 Copyright 2019 Thorlabs...

Страница 3: ...6 RS232 Operation 13 3 2 7 UART Operation 13 3 2 8 Reset Button 13 3 2 9 Baud Rate 13 3 2 10 Non Volatile Memory 13 3 2 11 Auxiliary I O Ports 14 3 2 12 Analog Output 14 3 2 13 Autonomous Operation to...

Страница 4: ...4 5 Return of Devices 21 4 6 Manufacturers Address 22 4 7 Certifications and Compliances 23 4 8 Warranty 23 4 9 Copyright and Exclusion of Reliability 24 4 10 Thorlabs Worldwide Contacts and WEEE Poli...

Страница 5: ...king forward to hearing from you Thorlabs GmbH Warning Sections marked by this symbol explain dangers that might result in personal injury or death Always read the associated information carefully bef...

Страница 6: ...interface for instrument con trol and to connect a power supply Further the PM102 can even be controlled via an external micro controller supplied by the customer For more information on the special f...

Страница 7: ...nd Operation SMA Output Connectors DA 15 Universal Connector RS232 Operation Analog Output Operation Autonomous Operation to Analog Output Sensitivity Configurable Analog Output UART Operation 2 Auxil...

Страница 8: ...ini B with locking screw 1 5m PM102A USB Cable Type A to Mini B with locking screw 1 5m PM102U USB Cable Type A to Mini B with locking screw 1 5m Optional Accessories To run the PM102x power meter add...

Страница 9: ...can be run with the Thorlabs OPM software Please see the require ments for the OPM on the website Sensor Requirements Thorlabs C Series Sensors The PM102x supports all Thorlabs C Series thermal power...

Страница 10: ...contents for completeness and test the PM102x mechanically and electrically Keep the container for storage or in order to return the product in case of future problems Verify that you have received t...

Страница 11: ...ument PC using the USB cable or a matching cable for the respective output port SMA or DA 15 For PM102A and PM102 please provide cables Connect a suitable sensor to the PM102x The sensor status LED in...

Страница 12: ...omatically recognize sensors lacking a DE 9 connector as for the case of the A and B series Please contact Thorlabs for an upgrade of old sensors with DE 9 connectors Note Custom Sensors To use sensor...

Страница 13: ...l power measure ments A Reset Button allows quick rebooting of the micro controller The PM102x is powered either via the USB interface or via the DA 15 interface A status LED indicates appropriate con...

Страница 14: ...data export The PM102 can also be adjus ted to accommodate micro controllers using 3 3 V 1 DA 15 Interface 2 Two Auxiliary I O ports 3 NTC input 4 Three Analog Output Ports 5 Autonomous Operation to A...

Страница 15: ...l Features 1 Autonomous Operation to analog output 2 Three Analog Output Ports POWER for the sensor signal X POS and Y POS for the posi tion of the signal in X and Y on thermal position sensors 3 Conf...

Страница 16: ...2019 Thorlabs 11 PM102x 3 2 4 PM102U PM102U Special Features Power Meter Control and Output via USB...

Страница 17: ...32 operation or UART operation the PM102 uses the DA 15 interface for control and output 3 2 6 RS232 Operation The PM102 can be operated with serial communication via RS232 Please use a RS232 standard...

Страница 18: ...to the in ternal non volatile memory 3 2 9 Baud Rate For PM102 the Baud rate can be adjusted from 9600 to 230400 Bit s This can be done either via the USB interface or via the serial interface using...

Страница 19: ...ands for custom made software The constant can be set in the Thorlabs software OPM under the Settings tab POWER AO Responsivity The signal of the port POWER is generated with a 16 bit DAC with an upda...

Страница 20: ...and data management software can be found in the software manual directly downloaded here Note Do not connect the PM102x prior to software installation Attention Exit all running applications on your...

Страница 21: ...ts POWER X POS and Y POS There are two configuration options for these three ports One for the port POWER and one for the two ports X POS and Y POS to configure the position in X and Y 3 4 Maintenance...

Страница 22: ...ent on Sensor and Settings Time Constant Correction Range 1 s 30 s Wavelength Correction Sensor Dependent nm V W Beam Area Setting Diameter 1 e or Rectangular x y Sensor Temperature Control Supported...

Страница 23: ...nector 5 V USB 2 0 Power Consumption 1 W Display Display Type None External PC Windows Application or Driver Set Display Screens Windows Application Software Provided GUI Refresh Rate 1000 Hz Dependen...

Страница 24: ...round 12 GND Ground 13 GND Ground 14 N C NTC Thermistor Input Measurement Range 0 1 kW to 100 kW 15 GND RS232 Signal Ground Connect to PC Terminal Ground PC DE 9 Pin5 4 3 Pin Assignment of the Sensor...

Страница 25: ...ables Attention The following statement applies to the products covered in this manual unless otherwise spe cified herein The statement for other products will appear in the accompanying documentation...

Страница 26: ...e meters of this unit since the electromagnetic field intensity may then exceed the maximum allowed disturbance values according to IEC 61326 1 This product has been tested and found to comply with th...

Страница 27: ...2019 Thorlabs 4 Appendix 22 4 7 Certifications and Compliances...

Страница 28: ...the technical data of the de scribed unit at any time 4 9 Copyright and Exclusion of Reliability Thorlabs has taken every possible care in preparing this document We however assume no li ability for t...

Страница 29: ...thorlabs com Thorlabs End of Life Policy WEEE Thorlabs verifies our compliance with the WEEE Waste Electrical and Electronic Equip ment directive of the European Community and the corresponding natio...

Страница 30: ...www thorlabs com...

Отзывы: