MTD415T Data Sheet Rev. 1.0
6 Programmers Reference
6.1 Nomenclature
Program messages are written
in inverted commas:
"A?"
"Lx"
(
x
- Parameter)
Response messages are written in brackets:
[-800 <LF>]
Data Format
Numerical value with sign in integer notation:
-220
or
16789432
.
Command /response Terminator:
Line Feed (
<LF>
)
6.2 Command Description
6.2.1 General Commands
Command
Explanation
Response Example
"m?"
Reads the version of hardware and software
[MTD415T FW0.6.8]
Command
Explanation
Response Example
"u?"
Reads the UUID (Universal Unique Identifier) of the MTD415T
[045F778655FDE5118ED499C9B4521485]
Command
Explanation
Response Example
"E?"
Reads the Error Register. Responses
see section
Error Register and Safety Bitmask
"c"
Resets the Error register
Note
The MTD415T has a non-volatile memory (flash) that stores the setting parameters. This
memory has a limited number of erase / write cycles. In order to protect the flash memory,
changes to setting parameters are not stored automatically. If you want to keep parameter
changes after power-down of the MTD415T, save them to the flash memory using the "
M
"
command. This command saves the T, W, L, d, G, O, P, I, D, C and S parameters at a time.
18