6.3 Beam Quality (M²) Measurement
The
Beam Quality
panel can be opened from the menu bar (Menu
Windows
) or by clicking
the
icon in the tool bar.
6.3.1 General
The M² value is an important measure of the beam quality. It is widely used in the laser industry
as a specification, and its method of measurement is defined in the ISO 11146 standard. It is
particularly useful to describe the degree of divergence and the focusability (minimum focus
diameter) of a real laser beam.
For more detailed information about Beam Quality, please see section
.
Beam Quality
Beam Propagation measurements according to the
standard
ISO 11146
describe the beam
quality using a single parameter, which is either the
Times-Diffraction-Limit Factor M
2
or the
Beam quality K = 1/M
2
(also known as
beam quality factor
or
beam propagation factor
).
Whereas the beam quality K is directly proportional to the quality level (K=1 optimal, decreasing
K stands for poorer quality), its reciprocal value M
2
(M
2
=1 optimal, increasing M for poorer
quality) is used more often.
Please do not confuse the beam quality
and times-diffraction-limit factor (M
2
=1).
Diffraction Limit
Depending on the wavelength
l
and beam divergence angle
q,
the theoretical limit for the mini-
mum beam waist diameter d
0
is called the
diffraction limit
. The beam waist cannot be de-
creased beyond this value.
M
2
expresses how close the diffraction limit of the analyzed beam is to the diffraction limit of an
ideal Gaussian beam. The beam parameter product d
0
*
q
describes that mathematically. For
beams of less quality, the product d
0
*
q
is increased by the factor M
2
.
2
0
0
d θ
M d θ
®
where d
0
is beam waist at the focus and
q
- the divergence angle.
M
2
is also known as
·
the ratio of the waist diameter d
0
of the measured beam to that of an ideal Gaussian
beam (TEM
00
) at the same divergence angle
q
.
·
the ratio of the divergence angle
q
of the measured beam to that of an ideal Gaussian
beam (TEM
00
) at the same waist diameter d
0
,
Lower beam quality results from laser imperfections like inhomogeneities which lead to appear-
ance of higher transverse modes.
140
Содержание BC207UV
Страница 2: ...Copyright 2022 Thorlabs Version Date 8 1 10 Jan 2022 ...
Страница 103: ...BC207 Series Chapter 6 Operation Instructions Rev 8 1 10 Jan 2022 Page 96 2022 Thorlabs ...
Страница 133: ...BC207 Series Chapter 6 Operation Instructions Rev 8 1 10 Jan 2022 Page 126 2022 Thorlabs ...
Страница 175: ...BC207 Series Chapter 10 Appendix Rev 8 1 10 Jan 2022 Page 168 2022 Thorlabs 10 11 Certifications and Compliances ...
Страница 181: ...www thorlabs com ...