2.3. 

Battery String Test 

 

The Battery String Test interface is as follows, and its Test operation method is similar to that of single cell Test, 

but the number of battery cells is increased, and before a String of batteries is tested, it can be tested continuously 

without operating the interface. Click the touch screen Start Test button to take the Test. 

 

2.3.1

 Custom standard parameters

 

Click the "Unit" button on the right side of "Battery Voltage", pop up the custom parameter dialog window, set the 

custom  parameters,  click  "Save".  After  saving,  you  can  see  that  the  "V"  of  the  button  changes  to  "Custom  x", 

indicating that the customization was successful. 

2.3.2

 Standard parameter selection

 

Click the button  on the right side of the "Battery Voltage" or "Standard Capacity" character to pop up the list of 

standard parameters, select  a  parameter according  to  the  needs,  click again to set the  parameters, and the  main 

interface updates the display. After clicking the "Save" button at the bottom of the "String Test" interface, the set 

parameters  will  be  saved  permanently  and  the  switch  will  not  be  lost  again,  and  there  is  no  need  to  set  the 

parameters repeatedly, which is convenient for the next test. 

Содержание DFT-6108

Страница 1: ...DFT 6108 Battery Analyzer Users Manual THOMSON TECH CO LTD www thomsontesting com...

Страница 2: ...n 9 2 6 1 File management 9 2 6 2 Language selection 9 2 6 3 Test the waveform 10 2 6 4 Software Version information 10 Chapter 3 Tester Main Parts Structure 12 3 1 Package List 12 3 2 DFT 6108 Host 1...

Страница 3: ...accurately test battery voltage current impedance ohmic resistance polarization resistance electrode double layer capacitance and other battery characteristic parameters which can be used for power co...

Страница 4: ...yze and manage the test data and can automatically generate corresponding reports The tester design is small light weight and easy to carry The tester program can be customized and updated according t...

Страница 5: ...1 0 rdg 6dgt Voltage 0 2 rdg 6dgt Capacitance 1 0 rdg 6dgt Working power supply 4000mAH rechargeable battery as full for 8 12 hours Memory capacity Stores 100 000 battery measurement data Display 4 3...

Страница 6: ...operation area When using the touch screen you can tap the menu task item directly and then tap the selected menu task item to execute If you are using a keyboard use the TAB key to toggle the selecte...

Страница 7: ...measured again if the test clip and the battery are not in good contact the measurement result is very different and at this time you can re measure again Click the Waveform button to view the battery...

Страница 8: ...meter dialog window set the custom parameters click Save After saving you can see that the V of the button changes to Custom x indicating that the customization was successful 2 3 2 Standard parameter...

Страница 9: ...a which can be opened and played back on the data records Dump U disks Delete and other operations while also formatting data records The interface for data management is shown in the following figure...

Страница 10: ...ftware through the U disk first put the upgraded software image file into the U disk Power Off then connect the U disk and the tester through the OTG cable press back key to boot and Power on and then...

Страница 11: ...ck key to return to the system setup interface 2 6 3 Test the waveform As an auxiliary function it can test 0 200V DC voltage of which is convenient for users auxiliary tests 2 6 4 Software Version in...

Страница 12: ...DFT 6108 Battery Analyzer Instruction Manual 11...

Страница 13: ...e List Battery Resistance Tester DFT 6108 Host 1pc Test clamp or Probe 1kit Host Charger 1kit U disk 1pc Data line 1pc Box 1pc User manual 1pc Packing list 1pc 3 2 DFT 6108 Host The host is used for f...

Страница 14: ...8 Battery Analyzer Instruction Manual 13 3 3 Testing Line Testing Line is either or of Testing Line clamp or Testing Line probe 3 4 Host Charger Host Charger is for Host inside battery charging Quanti...

Страница 15: ...DFT 6108 Battery Analyzer Instruction Manual 14 3 5 Data Line Data Line is for data communication from host to PC by U disk 3 6 Box Used to pack the main host and accessories Quantity 1pcs...

Страница 16: ...data 1 Boot screen data Logpic BMP Length X Width 272X480 2 Test result data File name AAACCCC DAT AAA is REM CCCCC is serial number Such as REM00035 DAT 4 3 Data management Under the Data Management...

Страница 17: ...ized and placed in the appropriate position in the chassis In order to keep the battery in optimal condition it is recommended to charge the battery regularly once a month 5 3 Tester s Battery mainten...

Страница 18: ...R The voltage of the measured battery V Voltage of a standard battery V Voltage ratio V Battery status judgment method R 1 15 the battery shows as excellent 1 15 R 1 5 the battery shows as good 1 5 R...

Страница 19: ...DFT 6108 Battery Analyzer Instruction Manual 18 6 3 Battery internal resistance list...

Страница 20: ...DFT 6108 Battery Analyzer Instruction Manual 19 Battery Data Management PC Software Manual THOMSON TECH CO LTD...

Страница 21: ...Software runtime environment 21 1 2 Software installation 21 1 3 Software uninstallation 21 1 4 Software running 21 1 5 Main Menu 22 Chapter 2 Data viewing and exporting to reports 23 2 1 Data Logger...

Страница 22: ...he following figure Click Done B PC Software installation Find the BTLStudioV1 3 4 on the U disk double click to run as shown in the following figure Click Next as shown below Select the corresponding...

Страница 23: ...DFT 6108 Battery Analyzer Instruction Manual 22 1 5 Main Menu...

Страница 24: ...exporting to reports 2 1 Data Logger monitoring For the monitoring of the battery Data Logger the Site must be added before monitoring and the establishment process is shown in the following figure A...

Страница 25: ...n 2 2 Data viewing and importing old report Import Data Click the Import Data icon button pop up the interface select a file you can choose a file to import the import success will show success failur...

Страница 26: ...the Save Report button in the lower right corner as shown below You can see the data report in excel or RTF format After the report export is completed the report will pop up in Sheet1 you can see the...

Страница 27: ...6108 Battery Analyzer Instruction Manual 26 2 4 Exit the software Click Close Report to exit the data report Click Close Recording to close the Data Logging page Click Exit System to close the softwa...