SCPI Command Set
[:SENSe]:FREQuency:SHIFt
This command sets or queries the frequency shift value. A negative shift value
corresponds to a left shifting.
This command is also used in additional to
the WSA down to 1Hz resolution.
Note:
Frequency shift mode is not available for some RFE modes of operation. Also,
when enabled, it would affect the data output format of some RFE modes. See
Radio RFE Modes and DSP Data Output Formats
).
Syntax
[:SENSe]:FREQuency:SHIFt <NRf [unit]>
[:SENSe]:FREQuency:SHIFt? [MAX | MIN]
Parameters
<Frequency [unit]>
Input Data Type
Float [character]
Allowable Values
-62.5 – 62.5MHz
Query Response
<integer>
Default I/O Unit
Hz
*RST State
0
Examples
:FREQ:SHIF -10.5 MHz
SENSE:FREQ:SHIFT 20000000.0
SENSe:FREQ:SHIFT?
FREQ:SHIFT? MAX
[:SENSe]:FREQuency:RESolution?
This command queries the frequency resolution value.
Syntax
[:SENSe]:FREQuency:RESolution?
Parameter
None
Query Response
<integer>
Default Output Unit
Hz
Example
:FREQ:RES?
[:SENSe]:LOCK:REFerence?
This command queries the lock status of the PLL reference clock in the digital card.
ThinkRF WSA5000 Wireless Signal Analyzer Programmer's Guide
67