7.2.5 Diagnostics commands
Command DA
<id>DA
Output of the optional measuring channels (5.4110.1x.xxx)
Description: Temperature/humidity:
30s-mean value (measurement every second)
Wind speed:
1s-mean value
Wind direction:
instantaneous value (measurement every second)
column Description
1-5 Temperature
[°C]
7-11
Relative air humidity [%r.h.]
13-16
Wind speed [m/s]
18-20
Wind direction [°]
If a sensor is unavailable or the measuring range exceeded the value of the
relevant column is “9”.
Example
reply:
„-01.6;040.3;02.6;090“
(-1.6 °C temperature; 40.3 % relative humidity,
2.6 m/s wind speed, 90 ° wind direction)
Value range:
-
Initial value:
-
Command DD
<id>DD
Request of the diagnostic data
Description:
Number Column Description
1
1-3
Interior temperature [°C]
2
5-6
Temperature of laser driver 0-80°C
3
8-11
Mean value laser current [1/100 mA]
4
13-16
Control voltage [mV] reference value: 4010
±
5
5
18-21
Optical control output [mV] / control variable
(500 … 6500)
6
23-25
Voltage sensor supply [1/10V]
**
7
27-29
Current pane heating laser head [mA]
**
8
31-33
Current pane heating receiver head [mA]
**
9
35-39
Ambient temperature [°C]
*
10
41-43
Voltage Heating supply [1/10 V]
*
11
45-48
Current heating housing [mA]
*
12 50-53
Current
heating heads [mA]
*
13 55-58
Current
heating carriers [mA]
*
32 - 66
021341/07/11