HK/HB/HG: switch on heating for 1 minute (0:Off 1:On) head(K), holder(B),
housing(G)
7.2.1 Communication commands
Command KY
<id>KY<para5>
Setting the command mode (0: normal, 1: configuration)
Description:
Two modes are available:
•
Normal mode (no setting of parameter possible)
•
Configuration mode (all commands allowed)
Value range:
0,1
Initial value:
0 (set to zero by power up, no storing)
Command BR
<id>BR<para5>
Setting the baud rate.
description:
The communication can be carried out with different baud rates and parities.
The setting range is from 1200baud to 115.2Kbaud, 7 or 8 data bits, no(N),
even(E), odd(O) parity, 1 or 2 stop bits. The following parameter are defined for
BR:
Parameter description:
1200 2400 4800 9600 19200
38400
57600
115.2K
8N1
2 3 4
5 *
6 7 8 9
7E1
10 11 12 13 14 15 16 17
8E1
18 19 20 21 22 23 24 25
8O1
26 27 28 29 30 31 32 33
7N1
34 35 36 37 38 39 40 41
7O1
42 43 44 45 46 47 48 49
8N2
50 51 52 53 54 55 56 57
7E2
58 59 60 61 62 63 64 65
8E2
66 67 68 69 70 71 72 73
8O2
74 75 76 77 78 79 80 81
7N2
82 83 84 85 86 87 88 89
7O2
90 91 92 93 94 95 96 97
Table 2: List of baud rates with telegram BR
* Setting by the factory
Value range:
2..97
Initial value:
00005