4.2.2 Connector for Precipitation Monitor without factory-connected Cable
For: 5.4103.10.000, 5.4103.11.000, 5.4103.10.112, see circuit diagram in chapter 4.
Process:
-
Remove 5 screws at the bottom side of the precipitation monitor.
-
Remove upper part (cover). The connecting terminals are the freely accessible.
-
Insert a respective prepared cable from the bottom through the cable glands situated
in the housing bottom, and connect it to the connecting terminals and shield
connector, acc. to circuit diagram.
-
Secure cable by cable glands.
-
Put on the upper part again and screw it evenly and tightly to the bottom.
Attetion:
Make sure, that the contact pins are not deformed when putting on the cover.
The fixing screws for the cover must be srewed down with a torsional of
1Nm to 2Nm
.
4.2.3 Electrical Mounting for Precipitation Sensor with Plug Connection
For: 5.4103.10.700, see circuit diagram in chapter 4.
4.2.3.1 Plug Mounting
Applies only to instruments with connection „plug“.
Coupling socket, type: Binder, Serial 423, EMC with cable clamp
Cable connection: without cable shield
1. Stringing parts on cable acc. to plan given
above.
2. Stripping cable sheath 20mm.
3. Cutting uncovered shield 20mm.
4. Stripping wire 5mm.
5. Soldering wire to the insert.
6. Positioning shield in cable clamp.
7. Screwing-on cable clamp.
8. Assembling remaining parts acc. to upper plan.
9. Tightening pull-relief of cable by screw-wrench
(SW16 und 17).
Kabelklemme
Kabelmantel
Kabellitzen