7.2 DIP-Switch Setting: 5.4103.10.000 / 700, 5.4103.11. 000,
5.4103.10. 112
DIP1- switch
Function: off delay
DIP2- switch
Function: drop incidences-filter
S 1
S 2
S 3
S 4
Time
(sec)
S 1
S 2
S 3
S 4
Drops
ON
OFF
OFF
OFF
25
ON
OFF
OFF
OFF
1
OFF
ON
OFF
OFF
50
OFF
ON
OFF
OFF
2
ON
ON
OFF
OFF
75
ON
ON
OFF
OFF
3
OFF
OFF
ON
OFF
100
OFF
OFF
ON
OFF
4
ON
OFF
ON
OFF
125
ON
OFF
ON
OFF
5
OFF
ON
ON
OFF
150
OFF
ON
ON
OFF
6
ON
ON
ON
OFF
175
ON
ON
ON
OFF
7
OFF
OFF
OFF
ON
200
OFF
OFF
OFF
ON
8
ON
OFF
OFF
ON
225
ON
OFF
OFF
ON
9
OFF
ON
OFF
ON
250
OFF
ON
OFF
ON
10
ON
ON
OFF
ON
275
ON
ON
OFF
ON
11
OFF
OFF
ON
ON
300
OFF
OFF
ON
ON
12
ON
OFF
ON
ON
325
ON
OFF
ON
ON
13
OFF
ON
ON
ON
350
OFF
ON
ON
ON
14
ON
ON
ON
ON
375
ON
ON
ON
ON
15
•
Grey marked squares = factory settings.
•
DIP – switch adjustment „OFF, OFF, OFF, OFF“: not defined.
7.3 DIP-Switch Setting: 5.4103.31.000
DIP1- switch
Function: off delay
DIP2- switch
Function: drop incidences-filter
S 1
S 2
S 3
S 4
Time
(sec)
S 1
S 2
S 3
S 4
Drops
ON
OFF
OFF
OFF
12,5
ON
OFF
OFF
OFF
1
OFF
ON
OFF
OFF
25
OFF
ON
OFF
OFF
2
ON
ON
OFF
OFF
37,5
ON
ON
OFF
OFF
3
OFF
OFF
ON
OFF
50
OFF
OFF
ON
OFF
4
ON
OFF
ON
OFF
62,5
ON
OFF
ON
OFF
5
OFF
ON
ON
OFF
75
OFF
ON
ON
OFF
6
ON
ON
ON
OFF
87,5
ON
ON
ON
OFF
7
OFF
OFF
OFF
ON
100
OFF
OFF
OFF
ON
8
ON
OFF
OFF
ON
112,5
ON
OFF
OFF
ON
9
OFF
ON
OFF
ON
125
OFF
ON
OFF
ON
10
ON
ON
OFF
ON
137,5
ON
ON
OFF
ON
11
OFF
OFF
ON
ON
150
OFF
OFF
ON
ON
12
ON
OFF
ON
ON
162,5
ON
OFF
ON
ON
13
OFF
ON
ON
ON
175
OFF
ON
ON
ON
14
ON
ON
ON
ON
187,5
ON
ON
ON
ON
15
•
Grey marked squares = factory settings.
•
DIP – switch adjustment „OFF, OFF, OFF, OFF“: not defined..