11.2.5 Command FB
<id>FB<parameter><CR>
Quick start mode
Access:
reading / writing
Description:
The command selects the quick start mode or the set mode is
queried.
Parameter type:
unsigned integer
Parameter:
0: Quick start mode off
1: Quick start mode on
Type return value:
unsigned integer
Return value:
see parameter
Value range:
0...1
Initial value:
1
11.2.6 Command FO
<id>FO<parameter><CR>
Frequency output
Access:
reading / writing
Description:
The command enables or disables the frequency output. When
no parameter is entered, the set value is queried.
Parameter type:
unsigned integer
Parameter:
0: deactivate frequency output
1: Output of measured frequency. The pulses from the internal
processor are acquired and re-output. The delivered frequency
is identical with the setting 7, with the difference that the pulse-
interval-ratio and the Jitter of the single current pulse are opti-
mized.
2: Output of the wind velocity by using the standard characteris-
tic curve (f
a
= WV* 10)
Example: WV=10,1m/s
fout=101Hz
3: Output of the wind velocity by using the latest calibration ta-
ble (f
a
= WG* 10)
Example: WV=10,1m/s
fout=101Hz
4: Output of the wind velocity by using the current calibration
table compensated concerning air pressure (f
a
= f
mess
* 10)
Example: WV=10,1m/s
fout=101Hz
5: reserved
6: reserved
7: Direct output of the measured pulse signals without applica-
tion of the calibration table or correction. In this mode the pulse
output behaves identically to 4.3351.X0.000
Type return value:
unsigned integer
Return value:
see parameter
Value range:
0...7
Initial value:
4